Company:Radiometer America Inc.
Date of Enforcement Report:4/11/2007
Class:ll
PRODUCT
ABL800 FLEX Series analyzer with FLEXQ module-equipped with software version 5.2 to 5.27, Recall # Z-0705-2007
REASON
In some situations pre-registered patient data is not reset when leaving the pre-registration mode in the software. This occurs when a Safe Pico blood sampler associated with a specific patient is pre-registered in the analyzer by scanning a barcode.
CODE
There are no specific lot codes. The problem is associated only with customers that use the ”Safe Pico” blood sampler in conjunction with the ABL800 analyzer.
RECALLING FIRM/MANUFACTURER
Recalling Firm: Radiometer America Inc., Westlake, OH, by letter dated January 16, 2007.
Manufacturer Address: Radiometer Medical ApS, Bronshoj, Denmark. Firm initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
1345 units
DISTRIBUTION
Nationwide and Internationally