Company: GE Healthcare LLC
Date of Enforcement Report 4/20/2011
Class ll:
PRODUCT
GE the CARESCAPE Monitor B850 includes both new and existing subsystems interconnected to form a high acuity, multi-parameter patient monitoring system. A typical configuration would be a CARESCAPE Monitor B850 host processing unit running the CARESCAPE Monitoring platform software, a display with integrated keyboard and a frame for the insertion of parameter measurement modules. A variety of options are available to the customer including additional displays, various input devices (keyboard, mouse, bar code reader), additional modules and frames and software options. The new subsystems include a host processing unit (CPU), a 15 inch LCD display with an integrated keypad, 19 inch LCD display with an integrated keypad and touch panel interface, a five slot parameter module frame, a seven slot parameter module frame, a cabled remote control, a cabled remote keypad, and the CARESCAPE Monitoring platform software. Some of these new major subsystems include non-patient contact accessory items (e.g. cables and mounting hardware). The CARESCAPE Monitor B850 interfaces to a variety of existing physiological parameter measurement modules. In addition, the CARESCAPE Monitor B850 interfaces to a variety of existing OEM medical devices via the existing network infrastructure. The CARESCAPE Monitor B850 is a multi-parameter high acuity patient monitor intended for use in multiple areas within a professional healthcare facility. The CARESCAPE Monitor B850 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE Monitor B850 system is indicated for monitoring of Hemodynamic (including ECG, ST Segment, Arrhythmia Detection, ECG Diagnostic Analysis and Measurement, Invasive Pressure, Noninvasive Blood Pressure, Pulse Oximetry, Cardiac Output, Temperature, Impedance Respiration and SpO2 (Mixed Venous Oxygen Saturation)), Airway Gases (Fi/Et C02, 02, N20 and Anesthetic Agent), Spirometry, Gas Exchange (02 Consumption (VO2), C02 production (VCO2), energy expenditure (BE), and respiratory quotient (RQ)) and neurophysiological (including electroencephalography (EEG), Entropy, Bispectral Index (BIS) and Neuromuscular Transmission (NMT) Monitoring) status. The CARESCAPE Monitor B850 provides alarms, trends, snapshots and events, and calculations and can be connected to displays, printers and recording devices. The CARESCAPE Monitor B850 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for bed to bed viewing and to data management software devices via a network. The CARESCAPE Monitor B850 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility. In addition to the healthcare practitioner, the CARESCAPE Monitor B850 is designed to provide configuration and troubleshooting capabilities to qualified service personnel. The CARESCAPE Monitor B850 is not intended for use during MRI. Recall #Z-1929-2011
REASON
GE Healthcare has become aware of potential safety issues associated with the CARESCAPE” Monitor B850 when Patient Data Module (PDM), Tram Acquisition Module (Tram), Patient Side Module (E-PSM or E-PSMP), E- MASIMO or E-(P)RE(S)TN modules or Bed-to-bed alarming feature is used. 1. When CARESCAPE Monitor B850 is used with the above listed modules, some SpO2 alarms may be delayed or missed. 2. When bed-to-bed alarm settings are done in the CARESCAPE Monitor B850 some alarm settings may be delayed or not changed as set by the clinician. 3. When Patient Data Module or TRAM module is used in transport, arrhythmia alarms may be missed under certain circumstances. 4. When CARESCAPE Monitor B850 is used with the Patient Data Module (PDM) with Masimo SpO2 technology, SpO2 Probe Off alarm may be missed.
RECALLING FIRM/MANUFACTURER
Recalling Firm: GE Healthcare, LLC, Waukesha, WI, by letter dated February 25, 2011.
Manufacturer: GE Healthcare Finland Oy, Helsinki, Finland. Firm initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
3,817 units
DISTRIBUTION
Nationwide and Internationally