Company:Bio-Rad Laboratories
Date of Enforcement Report 6/1/2011
Class ll:
PRODUCT
Software CD is labeled in part: “BLOOD VIRUS DIVISION***BIO-RAD EVOLIS System APF Version 4.4 Software Version 2.00 2009/05***” The EVOLIS Assay Protocol Files (APF) Software contains assay-specific instructions necessary for the EVOLIS Microplate processing system to process enzyme immunoassays on the EVOLIS instrument, perform data reduction and generate reports. Recall # Z-2201-2011
REASON
The EVOLIS Assay Protocol Files Software is configured in such a way that may result in a miscalculation of quantitative test results and/or an error in reporting positive/negative sample values..
RECALLING FIRM/MANUFACTURER
Recalling Firm: Bio-Rad Laboratories, Redmond, WA, by letter dated July 19, 2010, via FedEx.
Manufacturer: Bio Rad Laboratories, Marnes La Coquette, France. Firm initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
272 units
DISTRIBUTION
Nationwide