Company:Computerized Medical Systems Inc
Date of Enforcement Report 6/22/2011
Class ll:
PRODUCT
XiO Radiation Treatment Planning System, XiO Release 4.3.1 through 4.40.00. Used to create treatment plans for any cancer patient for who external beam radiation therapy or brachytherapy has been prescribed. Recall # Z-2567-2011
REASON
XiO Software: Under certain conditions, when the port shape is altered, dose is not recalculated. The port-change flag is not being set correctly when the port shape is regenerated. The dose distribution will not match the blocking displayed and the mis-match might not be obvious. Monitor Unit calculations will also be based on an incorrect dose distribution. This problem has existed since Release 4.3.1.
RECALLING FIRM/MANUFACTURER
Computerized Medical Systems Inc., Maryland Heights, MO, letter dated December 8, 2008. Firm initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
784 units
DISTRIBUTION
Nationwide and Internationally