Company:Philips Medical Systems
Date of Enforcement Report 4/25/12
Class ll:
PRODUCT
Brilliance Workspace IntelliSpace Portal, M/N 728269, The Brilliance Workspace Portal is a component of the Computed Tomography X-Ray System that is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories. System Code # 881001. Recall # Z-1337-2012
REASON
A number of non-conformances were logged for IntelliSpace Portal running software version 4.0 during in-house random verification testing. When loading a study to Review, after failing to load it through Add to Running Application the Analysis application is not being closed and different (1st loaded study) patient appears in analysis (CCA).
RECALLING FIRM/MANUFACTURER
Philips Medical Systems (Cleveland) Inc., Cleveland, OH, letter dated February 23, 2012. Firm initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
90 units
DISTRIBUTION
Nationwide and Internationally