Company:Animas Corp
Date of Enforcement Report 4/18/12
Class ll:
PRODUCT
OneTouch(R) Ping(R) Insulin Pump glucose management system. The device is intended to deliver insulin therapy. Part number 101420-00, 10142100, 101422-00, 101424-00, 101425-00, 100430-00, 100431-00, 100432-00, 100434-00, 100435-00, 101420-98, 101421-98, 101422-98, 101424-98, 101425-98, and 10050000. Recall # Z-1376-2012
REASON
There is an error in the pump software where the OneTouch(R) Ping(R) insulin pump did not allow the user to save changes to the time and date on the pump if those changes were made on February 29, 2012 (Leap Day). Users who did not attempt to change the time or date on their pump on February 29, 2012 are not affected by this issue.
RECALLING FIRM/MANUFACTURER
Animas Corp., West Chester, PA, by letters dated March 6, 2012. Firm initiated recall is ongoing.
DISTRIBUTION
Nationwide, Canada and Puerto Rico