Company:Intel-GE Care Innovations LLC.
Date of Enforcement Report 3/6/2013
Class ll:
PRODUCT
QuietCare-Networked product, Facility Server/model: QC101601; The QuietCare system is intended to detect and analyze residents’ movements.
Recall Number Z-0877-2013
REASON
Care Innovations detected a software anomaly: During the engineering investigation into the cause of why a Wander Alert was not generated, a defect in the implementation of the Wander and Night Motion ADLs was discovered.
RECALLING FIRM/MANUFACTURER
Intel-GE Care Innovations LLC, Rosevill, CA on 12/15/2012. Voluntary: Firm Initiated recall has been terminated
VOLUME OF PRODUCT IN COMMERCE
19 units
DISTRIBUTION
Nationwide