Company: Roche Diagnostics Operations, Inc.
Date of Enforcement Report: 12/4/2013
Class lI:
PRODUCT
ACCU-CHEK Aviva Combo Blood Glucose Meter (as a part of the ACCU-CHEK Aviva Combo System) The ACCU-CHEK Combo System is indicated for the treatment of insulin-requiring diabetes and for the quantitative measurement of glucose in fresh capillary whole blood from the finger. Recall Number Z-0385-2014
REASON
There is a software synchronization issue with the ACCU-CHEK Combo system. In rare cases, when the Manual Pump option is chosen on the ACCU-CHEK Aviva Combo meter, there is a possibility of receiving an incorrect bolus advice recommendation that may cause a temporary under delivery of insulin.
RECALLING FIRM/MANUFACTURER
Roche Diagnostics Operations, Inc., Indianapolis, IN on 10/25/2013. Voluntary: Firm Initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
6,595 units
DISTRIBUTION
Nationwide