Company:Elekta, Inc.
Date of Enforcement Report:3/19/2014
Class lI:
PRODUCT
XiO RPT System The XiO RTP System is used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed. The system will calculate and display, both on screen and in hard-copy, either two or three dimensional radiation dose distributions within a patient for a given treatment.
Recall Number Z-1197-2014
REASON
Dose and Monitor Unit values are not computed correctly when Elekta Motorized Wedges are used with Dynamic Conformal arcs.
RECALLING FIRM/MANUFACTURER
Elekta Inc., Atlanta GA. on 2/17/2014. Voluntary: Firm Initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
1776
DISTRIBUTION
Nationwide and Internationally
___________________________________