Company: Nellcor Puritan Bennett Inc. (dba Covidien LP)
Date of Enforcement Report:5/21/2014
Class lI:
PRODUCT
Recalled product is a component of the Newport Medical Instruments e360 Ventilator: single board computer, part number SBC2105A. Intended to provide invasive or noninvasive ventilator support and monitoring for infant, pediatric, and adult patients with respiratory failure or respiratory insufficiency.
Recall Number Z-1577-2014
REASON
Newport Medical Instruments, Inc. is conducting a voluntary field corrective action on specific Newport” e360 ventilators with a particular single board computer (SBC), in response to customer reports of the audible alarm not sounding when the ventilator is in an alarm state. The ventilator, as well as the visual and remote alarms, continue to operate normally.
RECALLING FIRM/MANUFACTURER
Nellcor Puritan Bennett Inc. (dba Covidien LP) Boulder, CO on 4/182014 Voluntary: Firm Initiated recall is ongoing
VOLUME OF PRODUCT IN COMMERCE
5,765
DISTRIBUTION
Nationwide and Internationally