Cl II adaPT insight (12C)

Company: on Beam Applications S.A.
Date of Enforcement Report 12/31/2014
Class lI:

PRODUCT

adaPT insight (12C), v1.3.2, PAT.109 (US), Treatment room 4: I2C is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy device. Recall Number Z-0836-2015

REASON
Two issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition workflow was linked to a wrong geometrical calibration file, resulting in an alignment offset for this specific mode of acquisition. 2. A gantry collision incident required the IBA service team to verify the gantry alignment and protons I x-ray beams colinearity, during which the colinearity parameters was changed. This was done for the Cone Beam Computed Tomography and kV/kV Dual Sources acquisition modes (the associated config ration file was modified), but not for the kV/kV Single Source acquisition mode (which requires to change another configuration file)..

RECALLING FIRM/MANUFACTURER
on Beam Applications S.A.A, Louvain La Neuve, BE on 11/7/2014. Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
1 unit

DISTRIBUTION
US Distribution in PA only.

___________________________________

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

62304 Software Training Course – February 23-25, 2021

IEC 62304 and Emerging Standards and FDA Expectations for Medical Device and Health IT Software – Virtual

This very popular 3-day course provides a clear understanding of applying IEC 62304 standard for medical device software and much more. The course compares and contrasts 62304 with FDA expectations and discusses approaches for alignment. In addition, participants will learn of other relevant standards and technical reports pertinent to medical device software, HealthIT, medical mobile apps, and Software as a Medical Device (SaMD) products (e.g., 82304, 80002-1, 14971, 80001-2-x, 62366).

Participants will gain practical advice and pragmatic experience with all types of medical software. Participants will leave with a clear understanding of how to effectively and efficiently integrate 62304 compliance into their software development lifecycle (SDLC).

Register (click):  EventZilla Registration Site

Need info?  Email us at training@softwarecpr.com

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.