Cl II Philips Healthcare Brilliance iCT xray

Company: Philips Healthcare
Date of Enforcement Report 6/15/2016
Class lI:

PRODUCT

Philips Healthcare Brilliance iCT Computed Tomography X-Ray System
Recall Number Z-1714-2016

REASON
Software Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of view; (4) DoseRight algorithm miscalculations leading to incorrect CTDI values; (5) scan length changes during subsequent axial results, (6) surview scan lengths near 135mm or 184mm

RECALLING FIRM/MANUFACTURER
Philips Healthcare, Andover, MA on 9/6/2015. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
423

DISTRIBUTION
Nationwidewide

___________________________________

About the author

Amy enjoys researching and writing about developments in medical technology and how that intersects with US law. She received her J.D. from the University of Florida Levin College of Law in 2020 and now works as a Regulatory Associate for SoftwareCPR®, a general-purpose regulatory consulting firm that is recognized globally for their expertise with standards and national regulations pertaining to medical device, mobile medical app, and HealthIT software.

Cybersecurity Review

Our cybersecurity experts are NESSUS Pro Licensed and can quickly remediate cybersecurity deficiencies with your medical device or digital health software.  Planning, requirements, validation, and submissions – we can assist with all.

Interested in having a conversation?  Email us to arrange a Zoom meeting or call us at +1 781-721-2921.

office@softwarecpr.com

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN), Canada, and Italy.