Company: The Metrix Company
Date of Enforcement Report 8/3/2016
Class lI:
PRODUCT
AUTOCOMP6 XP High Speed Compounder REF 58810
Recall Number Z-2258-2016
REASON
An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications.
RECALLING FIRM/MANUFACTURER
The Metrix Company on 6/23/2016 Voluntary: Firm Initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
93 devices
DISTRIBUTION
Nationwide and Internationally