Company: Elekta, Inc.
Date of Enforcement Report 8/31/2016
Class lI:
PRODUCT
Monaco RTP System Used to make treatment plans for patients with prescriptions for external beam radiation therapy.
Recall Number Z-2712-2016
REASON
In a specific workflow where contours are edited (enlarged, moved, copied, etc.) and then the contour is deleted on some slices and saved in the same session, it is possible that the deleted contours are still present which would not be the intent of the planner.
RECALLING FIRM/MANUFACTURER
Elekta inc., Atlanta, GA on 9/1/2016. Voluntary: Firm Initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
3,012 units
DISTRIBUTION
Nationwide and Internationally
___________________________________