Company: GE Healthcare It
Date of Enforcement Report 9/7/2016
Class lI:
PRODUCT
Centricity Laboratory Core Lab System 4.1 The Centricity Laboratory System is intended to be an information system designed to support the clinical and administrative activities associated with the provision and utilization of clinical laboratory services and facilities, e.g., the storing and delivering of analytical results. It is a specially designed data program application (software), which is supplied for installation in existing mainframe or decentralized computers or a computer network.
Recall Number Z-2671-2016
REASON
The Centricity Laboratory removes free-text notes on a master panel when the user deletes the slave panel. The deletion of a slave panel will (auto slash) the corresponding test (item) results on the master panel. Additionally, the displayed and HL7 results show neither the appropriate abnormal flag value nor the reference range values, when using large_num.g >7 to report numeric results.
RECALLING FIRM/MANUFACTURER
Ge Healthcare IT, Barrington, IL on 8/26/2016. Voluntary: Firm Initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
17
DISTRIBUTION
Nationwide and Internationally
___________________________________