Company:Varian Medical Systems, Inc.
Date of Enforcement Report 1/18/2017
Class lI:
PRODUCT
Eclipse Treatment Planning System version 13.MR2 [13.06.31 with Smart Segmentation Knowledge Based Contouring version 2.3 [2.3.12] Radiology: The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments.
Recall Number Z-0990-2017
REASON
Modifications in version 13.6MR2 for Contouring, SmartAdapt, and SmartSegmentation workspaces resulted in contours not being saved consistently in Eclipse. Treatment Planning System. The issue only occurs if certain conditions are fulfilled.
RECALLING FIRM/MANUFACTURER
Varian Medical Systems, Inc., Palo Alto, CA on 11/2/2016. Voluntary: Firm Initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
42 devices are affected
DISTRIBUTION
US Distribution to the states of : NJ, TN. OR and FL.