Company:Philips Electronics North America Corporation
Date of Enforcement Report 3/7/2018
Class lI:
PRODUCT
Intellivue iX Information Center Software – All PIIC iX Surveillance stations including: 866023 IntelliVue Info Center iX A.0 866117 PIIC Classic Upgrade 866389 IntelliVue Info Center iX B.0, C.0 867141 IntelliVue Info Center iX B.0 866025 IntelliVue Server IX A.0 866118 Database Server Upgrade A.0
Recall Number Z-0776-2018
REASON
Once a surveillance station is restarted on January 1, 2018 or later, the station will be unable to perform patient discharge and transfer operations. Any subsequent attempt to perform these operations will cause the station to restart, resulting in a short period of loss of monitoring at the Surveillance station during such restart. Until this issue can be corrected, users should avoid intentionally restarting their Surveillance stations in 2018.
RECALLING FIRM/MANUFACTURER
Philips Electronics North America Corporation, Andover, MA on 1/3/2018. Voluntary: Firm Initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
20357
DISTRIBUTION
Nationwide and Internationally