Recall – Exercise Testing Treadmill

Company: GE Medical Systems Ultrasound & Primary Care Diagnostics, LL
Date of Enforcement Report: 7/25/2018
Class II:

PRODUCT
T2100 Micro flex Drive Treadmill, powered Product Usage: Powered exercise equipment consist of powered devices intended for medical purposes, such as to redevelop muscles or restore motion to joints or for use as an adjunct treatment for obesity. Specifically the T210 Treadmill is intended for use in Exercise Testing, facilitating accurate blood pressure measurements and exercise testing within speed range of 0-13.5 miles per hour. Recall Number: Z-2467-2018

REASON
A performance issue with customer owned spare parts, T2100 Microflex drive (2026182-002 or 2026182-004), was not addressed with a previous safety correction. If these parts were installed from customer owned stock on the T2100 Treadmill, uncontrolled walking belt motion during a stress exercise test could occur.

RECALLING FIRM/MANUFACTURER
GE Medical Systems Ultrasound & Primary Care Diagnostics, LL, Milwaukee, WI on 4/2/2018. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
3,721 devices total

DISTRIBUTION
Nationwide and Internationally

62304 Software Training Course – February 23-25, 2021

IEC 62304 and Emerging Standards and FDA Expectations for Medical Device and Health IT Software – Virtual

This very popular 3-day course provides a clear understanding of applying IEC 62304 standard for medical device software and much more. The course compares and contrasts 62304 with FDA expectations and discusses approaches for alignment. In addition, participants will learn of other relevant standards and technical reports pertinent to medical device software, HealthIT, medical mobile apps, and Software as a Medical Device (SaMD) products (e.g., 82304, 80002-1, 14971, 80001-2-x, 62366).

Participants will gain practical advice and pragmatic experience with all types of medical software. Participants will leave with a clear understanding of how to effectively and efficiently integrate 62304 compliance into their software development lifecycle (SDLC).

Register (click):  EventZilla Registration Site

Need info?  Email us at training@softwarecpr.com

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.