Recall – Exercise Testing Treadmill

Company: GE Medical Systems Ultrasound & Primary Care Diagnostics, LL
Date of Enforcement Report: 7/25/2018
Class II:

PRODUCT
T2100 Micro flex Drive Treadmill, powered Product Usage: Powered exercise equipment consist of powered devices intended for medical purposes, such as to redevelop muscles or restore motion to joints or for use as an adjunct treatment for obesity. Specifically the T210 Treadmill is intended for use in Exercise Testing, facilitating accurate blood pressure measurements and exercise testing within speed range of 0-13.5 miles per hour. Recall Number: Z-2467-2018

REASON
A performance issue with customer owned spare parts, T2100 Microflex drive (2026182-002 or 2026182-004), was not addressed with a previous safety correction. If these parts were installed from customer owned stock on the T2100 Treadmill, uncontrolled walking belt motion during a stress exercise test could occur.

RECALLING FIRM/MANUFACTURER
GE Medical Systems Ultrasound & Primary Care Diagnostics, LL, Milwaukee, WI on 4/2/2018. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
3,721 devices total

DISTRIBUTION
Nationwide and Internationally

SoftwareCPR Training Courses:

IEC 62304 and other emerging standards for Medical Device and HealthIT Software

Our flagship course for preparing regulatory, quality, engineering, operations, and others for the activities and documentation expected for IEC 62304 conformance and for FDA expectations. The goal is to educate on the intent and purpose so that the participants are able to make informed decisions in the future.  Focus is not simply what the standard says, but what is meant and discuss examples and approaches one might implement to comply.  Special deep discount pricing available to FDA attendees and other regulators.

3-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Brian Pate

Next public offering:  TBD

Email training@softwarecpr.com to request a special pre-registration discount.  Limited number of pre-registration coupons.

Registration Link:

TBD

 


 

Being Agile & Yet Compliant (Public or Private)

Our SoftwareCPR unique approach to incorporating agile and lean engineering to your medical device software process training course is now open for scheduling!

  • Agile principles that align well with medical
  • Backlog management
  • Agile risk management
  • Incremental and iterative software development lifecycle management
  •  Frequent release management
  • And more!

2-days onsite (4 days virtual) with group exercises, quizzes, examples, Q&A.

Instructors: Mike Russell, Ron Baerg

Next public offering: March 7 & 28, 2024

Virtual via Zoom

Registration Link:

Register Now

 


 

Medical Device Cybersecurity (Public or Private)

This course takes a deep dive into the US FDA expectations for cybersecurity activities in the product development process with central focus on the cybersecurity risk analysis process. Overall approach will be tied to relevant standards and FDA guidance documentation. The course will follow the ISO 14971:2019 framework for overall structure but utilize IEC 62304, IEC 81001-5-1, and AAMI TIR57 for specific details regarding cybersecurity planning, risk characterization, threat modeling, and control strategies.

2-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Dr Peter Rech, 2nd instructor (optional)

Next public offering:  TBD

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.