Recall – Optical system displays unexpected behavior under certain conditions

Company: Leica Microsystems, Inc.
Date of Enforcement Report: 2/27/2019
Class II:

PRODUCT

PROVEO 8 Ophthalmic Microscope System optical instrument for improving the visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment.

Recall Number: Z-0838-2019

REASON

The observed issues can lead to unexpected interruptions of the optics carrier focus drive movement, intermittent stops of the X-Y coupling movement or sporadic failures in the proper on-screen display functionality of the HDR recording unit. Additionally, an anomaly in the device software may also lead to the unexpected behavior that when a user releases a handle or foot switch button this may not immediately end the actuated functionality. This behavior can only occur in unlikely workflows where a user changes a user profile or a mode change is initiated at the same time as a functional button is activated.

RECALLING FIRM/MANUFACTURER

Leica Microsystems, Inc. on 9/25/2017. Voluntary: Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE
142 units

DISTRIBUTION

Worldwide

62304 Software Training Course – February 23-25, 2021

IEC 62304 and Emerging Standards and FDA Expectations for Medical Device and Health IT Software – Virtual

This very popular 3-day course provides a clear understanding of applying IEC 62304 standard for medical device software and much more. The course compares and contrasts 62304 with FDA expectations and discusses approaches for alignment. In addition, participants will learn of other relevant standards and technical reports pertinent to medical device software, HealthIT, medical mobile apps, and Software as a Medical Device (SaMD) products (e.g., 82304, 80002-1, 14971, 80001-2-x, 62366).

Participants will gain practical advice and pragmatic experience with all types of medical software. Participants will leave with a clear understanding of how to effectively and efficiently integrate 62304 compliance into their software development lifecycle (SDLC).

Register (click):  EventZilla Registration Site

Need info?  Email us at training@softwarecpr.com

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.