Recall – Infusion Infusion pump software anomalies, Class I

Company: Fresenius Kabi
Date of Enforcement Report: 8/12/2019
Class I

PRODUCT

The Volumat MC Agilia Infusion System is an infusion pump used by health care professionals that delivers fluids including medications, blood, and blood products, into a patient’s body in controlled amounts. The pump administers fluids intravenously via an infusion tubing set and is intended for adults, pediatrics, and neonates. The infusion pump has a notification system based on different priorities of alarms (high, medium, and low) to notify clinicians of specific pump conditions that require their attention.

The optional Vigilant Agilia Drug Library is dose error reduction software (DERS) that is installed on a PC platform that communicates with the pump via a USB cable. It can be used to configure drug limits and pump settings according to clinical practice.

Recall Number: NA

REASON

When enabled, the “Keep Vein Open (KVO), End of Infusion” alarm triggers when the infusion is complete. This results in the therapeutic rate of a medication changing to a non-therapeutic rate, as defined by the KVO rate. KVO alarms should be high priority. If a health care provider fails to respond to a “KVO, End of Infusion” alarm for a critical medication, a delay in care and/or under-infusion of medication can take place, which may lead to death or serious injury because the patient will be receiving a non-therapeutic rate of medication.

Fresenius Kabi will change their  “Keep Vein Open (KVO), End of Infusion” alarm from a “low priority” to a “high priority” alarm to ensure the health care provider knows the infusion has completed.  There has been one death reported outside of the United States related to a norepinephrine infusion where a health care provider did not adjust the “Volume to be Infused” after changing the drug bag and then did not notice the pump’s “low priority” “KVO, End of Infusion” alarm. No other injuries have been reported in the rest of the world or in the United States.

Fresenius has also identified four software errors occurring with both the Volumat MC Agilia Infusion System and Vigilant Drug Library that may cause over-infusion or under-infusion of fluids/medications. These software errors could cause serious patient harm or death. There have been zero injuries and 14 complaints reported related to these software errors.

RECALLING FIRM/MANUFACTURER

Fresenius Kabi on 6/21/2019. Voluntary:  Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE

  • Volumat MC Agilia – 9,461
  • Vigilant Drug Library – 83

DISTRIBUTION

US

SoftwareCPR Training Courses:

Being Agile & Yet Compliant (Public)

Our SoftwareCPR unique approach to incorporating agile and lean engineering to your medical device software process training course is now open for registration!

  • Agile principles that align well with medical
  • Backlog management
  • Agile risk management
  • Incremental and iterative software development lifecycle management
  •  Frequent release management
  • And more!

3 days virtual (Zoom) with group exercises, quizzes, examples, Q&A.

Lead Instructor: Mike Russell

Next public offerings:

  • Americas: 11-13 February 2025
  • EU/Eastern Europe/Middle East/Africa/Atlantic/eastern South America: 18-20 February 2025
  • Southern Central Northeastern Pacific: 24-26 February 2025
Register using form at this link:     Agile Course Post Promo

 

IEC 62304 and other emerging standards for Medical Device and HealthIT Software

Our flagship course for preparing regulatory, quality, engineering, operations, and others for the activities and documentation expected for IEC 62304 conformance and for FDA expectations. The goal is to educate on the intent and purpose so that the participants are able to make informed decisions in the future.  Focus is not simply what the standard says, but what is meant and discuss examples and approaches one might implement to comply.  Special deep discount pricing available to FDA attendees and other regulators.

3-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Brian Pate

Next public offering:  TBD

Call or email now to schedule a private, in-house class. The fall schedule is filling up!

Email training@softwarecpr.com to request a special pre-registration discount.  Limited number of pre-registration coupons.

Registration Link:

TBD

 


 

Medical Device Cybersecurity (Public or Private)

This course takes a deep dive into the US FDA expectations for cybersecurity activities in the product development process with central focus on the cybersecurity risk analysis process. Overall approach will be tied to relevant standards and FDA guidance documentation. The course will follow the ISO 14971:2019 framework for overall structure but utilize IEC 62304, IEC 81001-5-1, and AAMI TIR57 for specific details regarding cybersecurity planning, risk characterization, threat modeling, and control strategies.

2-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Dr Peter Rech, 2nd instructor (optional)

Next public offering:  TBD

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.