Recall – Software fails to recognize new batteries

Company: Braemar Manufacturing, LLC
Date of Enforcement Report: 3/26/2020
Class II

PRODUCT

Braemar Manufacturing LLC DL900 Holter Monitor 1285 Corporate Center Dr. Suite 150 Eagan, MN 55121 USA Rx Only.

  • Recall Number: Z-1459-2020

REASON

An issue with the DL900 and DL950 Series Holter Recorder was recognized on January 1, 2020. This issue can prevent the recorder from operating normally. Braemar confirmed that beginning on January 1, 2020 if a AAA battery is inserted in the recorder and a user attempts to start it, the recorder will display Error: 602 and fail to function for a new patient study. There are no actions that a clinical user can take to clear this error. Even if this error code is displayed, existing ECG recordings stored on the device can still be downloaded via the standard Holter Application Software. Error 602 may be able to be cleared by technical or engineering staff at the customer site.

RECALLING FIRM/MANUFACTURER

Braemar Manufacturing, LLC on 1/22/2020. Voluntary:  Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE

10582

DISTRIBUTION

US Nationwide and International

SoftwareCPR Training Courses:

Being Agile & Yet Compliant (Public)

Our SoftwareCPR unique approach to incorporating agile and lean engineering to your medical device software process training course is now open for registration!

  • Agile principles that align well with medical
  • Backlog management
  • Agile risk management
  • Incremental and iterative software development lifecycle management
  •  Frequent release management
  • And more!

3 days virtual (Zoom) with group exercises, quizzes, examples, Q&A.

Lead Instructor: Mike Russell

Next public offering: Dec 3, 4, & 5, 2024 – 12:00 pm to 5:00 pm CET

Register Now


 

IEC 62304 and other emerging standards for Medical Device and HealthIT Software

Our flagship course for preparing regulatory, quality, engineering, operations, and others for the activities and documentation expected for IEC 62304 conformance and for FDA expectations. The goal is to educate on the intent and purpose so that the participants are able to make informed decisions in the future.  Focus is not simply what the standard says, but what is meant and discuss examples and approaches one might implement to comply.  Special deep discount pricing available to FDA attendees and other regulators.

3-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Brian Pate

Next public offering:  TBD

Call or email now to schedule a private, in-house class. The fall schedule is filling up!

Email training@softwarecpr.com to request a special pre-registration discount.  Limited number of pre-registration coupons.

Registration Link:

TBD

 


 

Medical Device Cybersecurity (Public or Private)

This course takes a deep dive into the US FDA expectations for cybersecurity activities in the product development process with central focus on the cybersecurity risk analysis process. Overall approach will be tied to relevant standards and FDA guidance documentation. The course will follow the ISO 14971:2019 framework for overall structure but utilize IEC 62304, IEC 81001-5-1, and AAMI TIR57 for specific details regarding cybersecurity planning, risk characterization, threat modeling, and control strategies.

2-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Dr Peter Rech, 2nd instructor (optional)

Next public offering:  TBD

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.