Recall – SOUP update leads to field failure

Company: Beckman Coulter Inc.
Date of Enforcement Report: 4/15/2020
Class II

PRODUCT

LabPro Data Management System, UDI: 15099590676872 – Product Usage: The LabPro Data Management System is a Microsoft Windows based software program and is intended to manage both microbial identification (ID) and antimicrobial agent susceptibility testing (AST) data generated from MicroScan instruments or manually entered microbiology test results, for use by trained laboratory personnel.

  • Recall Number: Z-1673-2020

REASON

A security-only update was released via remote diagnostics on January 14, 2020 to customers with Windows 7 operation system. Shortly after the releases some customers reported that Data Management System computers would not restart. If the computer is unable to restart, then the system cannot download data generated from MicroScan instruments or manually entered microbiology test results, resulting in a potential of delayed results.

RECALLING FIRM/MANUFACTURER

Beckman Coulter Inc. on 1/17/2020. Voluntary:  Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE

177 systems

DISTRIBUTION

US Nationwide and International

Cybersecurity Review

Our cybersecurity experts are NESSUS Pro Licensed and can quickly remediate cybersecurity deficiencies with your medical device or digital health software.  Planning, requirements, validation, and submissions – we can assist with all.

Interested in having a conversation?  Email us to arrange a Zoom meeting or call us at +1 781-721-2921.

office@softwarecpr.com

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN), Canada, and Italy.