Recall – Microscope eyepiece movement does not stop when button released

Company: Haag-Streit USA Inc
Date of Enforcement Report: 12/30/2020
Class II

PRODUCT

Haag Streit Surgical Floor stands FS 2-11 (Ref no. 615H511) and FS 2-15 (Ref no. 615H515) in combination with the operating microscopes HS Hi-R NEO 900 (Ref no. 657 820) and HS Hi-R NEO 900A (Ref no. 657 821), with software REF 615 588 versions 2.0 to 3.3 – Product Usage: use in ophthalmology and in other disciplines with vertical direction of view.

Recall Number: Z-0697-2021

REASON

Software error -Software REF 615 588 versions 2.0 to 3.3, movement of the focusing is triggered by pressing the button of the footswitch it might occur that the movement does not stop when releasing the button, result in potential risk to the patient’s eye.

RECALLING FIRM/MANUFACTURER

Haag-Streit USA Inc on 11/20/2020. Voluntary:  Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE

3 units US

DISTRIBUTION

U.S. Nationwide

62304 Software Training Course – February 23-25, 2021

IEC 62304 and Emerging Standards and FDA Expectations for Medical Device and Health IT Software – Virtual

This very popular 3-day course provides a clear understanding of applying IEC 62304 standard for medical device software and much more. The course compares and contrasts 62304 with FDA expectations and discusses approaches for alignment. In addition, participants will learn of other relevant standards and technical reports pertinent to medical device software, HealthIT, medical mobile apps, and Software as a Medical Device (SaMD) products (e.g., 82304, 80002-1, 14971, 80001-2-x, 62366).

Participants will gain practical advice and pragmatic experience with all types of medical software. Participants will leave with a clear understanding of how to effectively and efficiently integrate 62304 compliance into their software development lifecycle (SDLC).

Register (click):  EventZilla Registration Site

Need info?  Email us at training@softwarecpr.com

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.