Recall – Medical archive system allowed non-unique patient IDs

Company: Topcon Medical Systems, Inc.
Date of Enforcement Report: 2/10/2021
Class II

PRODUCT

Harmony Referral System (aka Harmony RS)/ Medical Device Data System – Product Usage: intended to be used by trained healthcare professionals and clinicians to access archived multimedia data, forms and patient information.

Recall Number: Z-0952-2021

REASON

Harmony RS integrations with Topcon equipment, TRC NW-400 and the Signal Camera, allowed user input of non-unique patient IDs that caused multiple patient records and images to be combined under the first generated record.

RECALLING FIRM/MANUFACTURER

Topcon Medical Systems, Inc. on 12/18/2020. Voluntary:  Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE

13 units

DISTRIBUTION

U.S. Nationwide

Cybersecurity Review

Our cybersecurity experts can quickly remediate cybersecurity deficiencies with your medical device or digital health software.  Planning, requirements, validation, and submissions – we can assist with all.

Interested in having a conversation?  Email us to arrange a Zoom meeting or call us at +1 781-721-2921.

office@softwarecpr.com

Corporate Office

15148 Springview St
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN) and Italy.