Recall – HVAD system fails to restart after being stopped, CL I

Company: Medtronic
Date of Enforcement Report: 3/1/2021
Class I

PRODUCT

HVAD Pump Implant Kits for the HeartWare HVAD System

  • PUMP 1103
  • PUMP 1104
  • PUMP 1104JP

The HeartWare Ventricular Assist Device (HVAD) Pump Implant Kit is part of the HeartWare HVAD System, which is used to help the heart continue to pump blood to the rest of the body. The HVAD system is used as a bridge to cardiac transplants in patients who are at risk of death from end-stage left ventricular heart failure, for heart tissue recovery, or as destination therapy (DT) in patients where new transplants are not planned.

Recall Number: Z-0946-2021

REASON

Medtronic is recalling the HVAD Pump Implant Kit because the device may fail to initially start, restart, or have a delay in restarting after the pump was stopped. These delays or failures to start or restart have occurred during preimplant testing, during the implant, or in a variety of post-implant situations. If the device has delays or fails to start or restart, this could cause serious patient harm including a heart attack, worsening heart failure, the need for additional procedures and hospitalizations, or death.

There have been 29 complaints about this device issue, which include 19 serious injuries and 8 cases of patients who had a life-threatening event but recovered without long term effects. Two deaths have been reported.

RECALLING FIRM/MANUFACTURER

Medtronic on 12/7/2020. Voluntary:  Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE

386 units

DISTRIBUTION

U.S. Nationwide and International

SoftwareCPR Training Courses:

Being Agile & Yet Compliant (Public)

Our SoftwareCPR unique approach to incorporating agile and lean engineering to your medical device software process training course is now open for registration!

  • Agile principles that align well with medical
  • Backlog management
  • Agile risk management
  • Incremental and iterative software development lifecycle management
  •  Frequent release management
  • And more!

3 days virtual (Zoom) with group exercises, quizzes, examples, Q&A.

Lead Instructor: Mike Russell

Next public offering: Dec 3, 4, & 5, 2024 – 12:00 pm to 5:00 pm CET

Register Now


 

IEC 62304 and other emerging standards for Medical Device and HealthIT Software

Our flagship course for preparing regulatory, quality, engineering, operations, and others for the activities and documentation expected for IEC 62304 conformance and for FDA expectations. The goal is to educate on the intent and purpose so that the participants are able to make informed decisions in the future.  Focus is not simply what the standard says, but what is meant and discuss examples and approaches one might implement to comply.  Special deep discount pricing available to FDA attendees and other regulators.

3-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Brian Pate

Next public offering:  TBD

Call or email now to schedule a private, in-house class. The fall schedule is filling up!

Email training@softwarecpr.com to request a special pre-registration discount.  Limited number of pre-registration coupons.

Registration Link:

TBD

 


 

Medical Device Cybersecurity (Public or Private)

This course takes a deep dive into the US FDA expectations for cybersecurity activities in the product development process with central focus on the cybersecurity risk analysis process. Overall approach will be tied to relevant standards and FDA guidance documentation. The course will follow the ISO 14971:2019 framework for overall structure but utilize IEC 62304, IEC 81001-5-1, and AAMI TIR57 for specific details regarding cybersecurity planning, risk characterization, threat modeling, and control strategies.

2-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Dr Peter Rech, 2nd instructor (optional)

Next public offering:  TBD

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.