Recall – Inadvertent restart of ventilator

Company: Draeger Medical, Inc.
Date of Enforcement Report: 10/20/2021
Class II

PRODUCT

Evita V500 Ventilator with Installed CO2 Measurement Option-Intended for the ventilation of adult, pediatric and neonatal patients.

  • Catalog Number: 8416400

Recall Number: Z-0108-2022

REASON

Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds.

RECALLING FIRM/MANUFACTURER

Draeger Medical, Inc. on 9/10/2021. Voluntary:  Firm Initiated recall is ongoing.

VOLUME OF PRODUCT IN COMMERCE

6,797 units

DISTRIBUTION

U.S. Nationwide

Cybersecurity Review

Our cybersecurity experts are NESSUS Pro Licensed and can quickly remediate cybersecurity deficiencies with your medical device or digital health software.  Planning, requirements, validation, and submissions – we can assist with all.

Interested in having a conversation?  Email us to arrange a Zoom meeting or call us at +1 781-721-2921.

office@softwarecpr.com