Company: Fisher & Paykel Healthcare, Ltd.
Date of Enforcement Report: 9/25/2024
Class II
PRODUCT
PT301US Airvo 3 Respiratory Support Device with software version 1.2.0 and/or 1.5.1 The Airvo 3 is intended to provide high flow warmed and humidified respiratory gases for administration to spontaneously breathing infant, child, adolescent and adult patients in hospitals and sub-acute facilities.
Recall Number:
Z-3262-2024
REASON:
Due to a software issue, affected devices that are set up with High Pressure Oxygen (HPO), if the flow alignment alarm occurs, the device will deliver room air only. If this happens, a patient my experience oxygen desaturation that could lead to hypoxia.
RECALLING FIRM/MANUFACTURER:
Fisher & Paykel Healthcare, Ltd.
VOLUME OF PRODUCT IN COMMERCE:
944 units
DISTRIBUTION:
U.S.: AZ, CA, CO, CT, DC, FL, GA, IA, IL, KY, MA, MD, MI, MN, MO, MT, ND, NH, NJ, NV, NY, OH, OR, PA, SD, TX, UT, VA and WA. O.U.S.: Not provided