Company:Hocoma AG,br>Date of Enforcement Report 1/27/2010
Class:ll
PRODUCT
Hocoma Lokomat-System with Pediatric Module Software Versions: V4.30, V4.31, V4.31a, V4.31b, V4.31c, V4.32, V4.32a, V5.0x. Powered exercise equipment used to apply automated locomoter training for hemiplegic and paraplegic patients. Recall # Z-0590-2010
REASON
Changes to the stored “sensitivity” values may occur and leads to higher limits in the force and trajectory deviation.p>
RECALLING FIRM/MANUFACTURER
Hocoma AG, Zuerich, Switzerland, by letter dated May 27, 2009. Firm initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
7 units
DISTRIBUTION
Nationwide