Company:I-Flow Corporation
Date of Enforcement Report 8/31/12
Class l:
PRODUCT
ON-Q Pump with ONDEMAND Bolus Button.
The I-Flow ON-Q pump with ONDEMAND bolus button was manufactured from October 18, 2002 through April 30, 2012 and distributed from May 3, 2010 through May 1, 2012
Use: The On-Q pump with ONDEMAND bolus button is used for continuous and intermittent delivery of medicines (such as local anesthetics or narcotics) to or around surgical wound sites and/or to nearby nerves for pre-operative, during the procedure/surgery (perioperative), and for post-operative regional anesthetic and pain management.
REASON
he firm determined that the ON-Q pumps with ONDEMAND bolus button may not lock in the down position when depressed and/or the orange bolus refill indicator may stay in the lowest most position. When this occurs, the patient may receive continuous infusion at a rate greater than expected. As a result, this product may cause serious adverse health consequences, including death.
Recalling Firm:
I-Flow Corporation,Lake Forest, California
FDA District: Los Angeles