Company: Mindray DS USA, Inc. dba Mindray North America
Date of Enforcement Report 2/10/2016
Class lI:
PRODUCT
Panorama Patient Monitoring Network. Software Kit 0020-00-0205-19A, View Station Software Kit 0020-00-0206-16A, Work Station Software Kit 0020-00-0207-16A, Software Kit Upgrade TECH-00-910. The Panaroma Patient Monitoring Network is intended for use in a fixed location in the healthcare facility setting as a central viewing station. It is not intended to be directly connected to patient at any time or installed in a patient’s vicinity.
Recall Number Z-0701-2016
REASON
An issue with the Panorama Central Station may cause the system to spontaneously restart. This may occur after about 49 days of continuous operation. On restart, monitoring will be lost for about 2.5 minutes. Settings and configurations are maintained..
RECALLING FIRM/MANUFACTURER
Mindray DS USA, Inc. dba Mindray North America, Mahwah, NJ on 12/15/2015. Voluntary: Firm Initiated recall is ongoing.
VOLUME OF PRODUCT IN COMMERCE
63 products
DISTRIBUTION
Nationwide