Company:Philips Ultrasound, Inc.
Date of Enforcement Report: 7/30/2014
Class lI:
PRODUCT
QLAB 10 Core Module PN 453561704771. QLAB Quantification Software is a software application package designed to view and quantify image data acquired on Philips Medical Systems ultrasound products. Recall Number Z-2109-2014
REASON
Philips Healthcare has discovered a problem in the Philips Ultrasound QLAB version 10.0 and 10.1 that could result in the erroneous calculation of Ejection Fraction (EF) from the results of a2DQ and aCMQ patient exams.
RECALLING FIRM/MANUFACTURER
Philips Ultrasound, Inc., Bothell WA on 7/9/2014 Voluntary: Firm Initiated recall is ongoing
VOLUME OF PRODUCT IN COMMERCE
84
DISTRIBUTION
Nationwide