Chronimed 3/2/99 Glucose test strips and meters 5. Failure to establish and maintain plans that describe or reference the design and development activities and define responsibility for implementation as required by 21 CFR 820.30(b). For example, there is no plan for the software validation for Version 104. 6. Failure to ensure that the design input...
Read More
All IMMULITE 2000 Automated Immunoassay Analyzers with SoftwareVersion 1.1, a clinical automated immunoassay analyzer intended for in-vitro diagnostic use for detection of analytates in human specimens. Recall #Z-909-9. REASON If the initial counts of the reaction tubes are exactly 10,000 cps, this leads to erroneously low calculated results. CODE All serial numbers in commercial distribution...
Read More
Abbott Cell-Dyn 3000 Series Hematology Systems, Cell-Dyn Reticulocyte Reagent, when used in combination with Cell-Dyn 3500 reticulocyte software enables a whole blood specimen to be analyzed for reticulocytes. The Reticulocyte Reagent is intended for in vitro diagnostic use (outside the body). Recall #Z-715-9. REASON These lots may exhibit elevated Reticulocyte results. In some cases, the...
Read More
Compex S.A. 2/22/99 Muscle and nerve stimulation devices 2. Failure to validate computer software for its intended use according to an established protocol, when computers or automated data processing systems are used as part of production or the quality system, as required by 21 CFR 820.70(i). For example, your firm failed to validate the software...
Read More
2/16/99 Class II 128XP Acuson 637 units. Rev 27 software with OB calculations options are affected. Potential error condition related to OB reports printed after diagnosis. Z-787-9
Read More
Solar 7000 and Solar 8000 Patient Monitors with Version 5D Software, used to display physiological data from modules which monitor the patient for ECG, blood pressure, cardiac output, respiration, pulseoximetry, etc. Recall #Z-727/728-9. REASON A software deviation can affect the alarm default settings, which may result in incorrect or missed alarm notification. CODE Solar 7000/8000...
Read More
Diehl Inc. Food Processing 2/12/99 The chart recorder and the computer display on the xxxx units did not correspond to the mercury-in-glass (MIG) thermometer. The chart recorder or computer display may show a reading up to 3 degrees F higher than the MIG thermometer. SoftwareCPR keywords: Production software, validation, electronic records
Read More
Inadequate process validation; failure to validate the test methods used to analyze raw materials and the finished product; failure to institute process qualification for the laboratory instruments used to perform in-process and finished product testing; failure to qualify or validate the firm’s computer software programs; and failure to perform investigations into out-of-specification weight and diameter...
Read More
10-Feb -99 Precision Therapy International. 190 units. Class II. 3-D Imgcomp Utility User Manual Update. Version 2.72b & 3.07d. A design error may leadto a serious miscalculation of radiation dose. Z-1223-9
Read More
2/9/99 Class II Bayer AADVIA120 Hematology System 920 units. The device may give erroneous reslts because of the valve gasket deterioration in the Baso channel check valve, and the software designed will not flag significant discrepancies in the results between the two methodologies at extremely low counts.Z-869/870-9
Read More
2/5/99 Class I recall Marquette Electronics ECG, blood pressure, cariac output, respiraion, pulse oximetry Monitors. 607 new units and 97 update kits. A software deviation can affect alarm default settings, which may result in incorrect or missed alarm notification. Z-727/728-9
Read More
2/1/99 Abbott Alcyon 300 Analyzer (IVD) Possible mixup of patient results. 631 units. Z-617-9
Read More
National Biomedical Computer System, Versions 1.1 to 1.3.2. Recall #B-861-9. REASON Y2K defect in the computer software incorrectly calculates the end date of a temporary deferral to be 1900 instead of 2000, and removes the deferral from the database. CODE Software versions 1.1 to 1.3.2. MANUFACTURER American Red Cross Blood Services, Arlington, Virginia. RECALLED BY...
Read More
Four Recalls of Picker International Nuclear Cameras and imaging systems12/15/98 Prism XPVT 63 units. Unexpected camera movement which requires the operator to power down and reboot with delay in diagnosis. Z-643/650-9 11/18/98 Model 3000 325 units. Unintended motion of gantry during setup. Z-641/642-9 11/17/98 ModelFX AND fx 800 WITH VERSION 8.3.11 software. 70 units. Software...
Read More
Knoll Pharmaceutical 12/14/98 Your firm’s equipment qualifications for the Amsco SG-116 autoclave and the computer control system for the 60″ film coating pan were inadequate. SoftwareCPR keywords: software validation, CDER, Drugs, Pharma
Read More
12/10/98 Datalink Management System Software Versions 4.1-4.2 Beckman Coulter 248 units. The devices may allow the test results from an associated clinical instrument to be placed under an incorrect sample ID. Z-829-9
Read More
Soft Computer Consultants 12/4/98 Blood Bank Computer SoftwareDESIGN CONTROLS 1) Failure to establish and maintain plans that describe or reference the design and development activities, and define responsibility for implementation including the plans that describe the interfaces with different groups that provide input to the design and development process [21 CFR 820.30(b)]. 2) Failure to...
Read More
PRODUCTVersion 6A and prior versions of software for Marquette Coherent Digital Telemetry (CDT) LAN Monitoring Systems, a patient monitoring system designed to collect and transmit ECG and other physiological data from ambulatory patients, without the patient being physically connected to a display device. Recall #Z-774-0.
Read More
Ultramedic Ltd. 11/25/98 Infusion Pump Analyzers 6. Failure to have an established protocol to validate computer software for its intended use when computers or automated data processing systems are used as part of production or the quality system, as required by 21 CFR 820.70(i). For example, there is no validation protocol for the validation study...
Read More
11/24/98 HMT Medical Technologies Switzerland. Lithotripsy instruments Software versions 3.0 and 1.2. 17 units. Software upgrade yielding error messages and complete lock-up of instruments. Z-680/681
Read More
North American Science Assoc 11/24/98 Contract Testing Laboratory Failure to maintain adequate controls over computers and related systems to assure that changes in the master production and control records or other records are instituted only by authorized personnel [21 CFR 211.68(b)]. For example: (1) there is no current listing of individuals who have access to...
Read More
North American Science Contract Test Laboratory 11/24/98 Failure to maintain adequate controls over computers and related systems to assure that changes in the master production and control records or other records are instituted only by authorized personnel [21 CFR 211.68(b)]. For example: (1) there is no current listing of individuals who have access to the...
Read More
11/2/98 ACS Centaur Automated Chemiluminescence System 244 units with Software Version 1.2 The device may go into premature ready state with bleach still remaining in the analyzeer lines due to a software error. Z-499-9.
Read More
Allergy Laboratories 8/14/98 8. There are no controls for computer generated labeling used for product inserts. It was noted during the inspection that finished HealthTest HIV 1 + 2 and HBsAg test kits were relabeled from “For in vitro use only” to “For Research Use Only”. SoftwareCPR keywords: software validation, CDRH, devices
Read More
ICS Medical 10/29/98 Diagnostic opthalmic devices 1. Failure to develop, monitor, and control the software development processes to ensure that the software used in _____ conforms to its original design or any approved changes. 2. Failure to review the software development process to identify actions needed to prevent the recurrence of software nonconformances. 3. Failure...
Read More
Villa Sistemi Medicali, S.p.A. 9/30/98 Dental x-ray systems Failure to maintain Device Master Records (DMR) that includes or refers to the location of: device specifications including component specifications, software specifications; production process specifications; quality assurance procedures; and labeling specifications as required by 21 CFR 820.181(a)(b)(c) and (d). For example, the firm has no Device Master...
Read More
Software for the Abbott AxSYM System, a fully automated immunoanalyzer. Recall #Z-331-9. REASON Extra assay results from a previous run can appear with new sample results, as well as error messages, due to software defect in the internal result database management. CODE All software versions below 3.03. MANUFACTURER Abbott Laboratories, Irving, Texas. RECALLED BY Abbott...
Read More
Review of the _____ which consisted of the software and hardware components necessary to _____ DHRs and maintain them as electronic records, revealed the following:. Changes to processes were not always reviewed and evaluated or revalidated, where appropriate, and documented [21 CFR 820.75(c)], in that the software validation dated _____ was completed for software _____...
Read More
9/4/98 Nera Electronics/Baxter Healthcare Flo Gard 8200 Volumetric Infusion Pump: Class II 13 units. Board assemblies had the wrong software version installed on them. The boards have version 2.09 instead of version 2.13. Recall #=-Z-777-8.
Read More
10/1/98 Class II Quinton Q-Cath Hemodynamic Analysis System 81 units. Units with version 4.5 software. Device has potential to display previous patient’s blood pressure. Z-855/859-9
Read More
GE Advantx Vascular Imaging Systems, angiographic X-ray systems: a) GE Advantx Vascular Imaging Systems (DX and DLX) with Stenosis Analysis Software and; b) Cardiac Review Stations (CRS) with Quantitative Analysis Software. Safety Alert #N-002/003-9. REASON When procedures in the Operator Manual for stenosis sizing are not followed each time a time measurement is taken, serious...
Read More
Blood Bank Control Systems. Recall #B-020-0. REASON Computer software contains programming errors that could potentially result in the release of a unit that has not been tested for antibodies. CODE Software versions 4.0 and greater. MANUFACTURER Blood Bank Computer Systems, Kent, Washington. RECALLED BY Manufacturer, by letter dated September 1, 1998. Firm-initiated field correction complete....
Read More
Bernafon-MAICO Inc. 5/06/97 “Computer Validation The inspection found that the xxxx Computer System, used to add bulk quantities of solvents to the manufacturing processes (from a tank farm) for Ticlopidine and Bromocryptine Mesylate drug substances has not been validated (FDA-483 Item #16). The response to FDA-483 Item #16 is inadequate. Your firm’s commitment is to...
Read More
Failure to validate and document significant manufacturing processes and quality assurance tests to assure specific requirements are met, e.g., robot and manual welding processes, and software used to program the chip in the control device of the automatic air mattress. Your firm’s response dated July 30, 1998 and signed by Edward A. Kroll, Director, TQM...
Read More
Sabratek Corp. 8/19/98 Infusion pumps 1. Failure to identify the actions needed to correct and prevent recurrence of non-conforming product. For example, it was determined that the _____ had a software problem which prevented the infusion rate from exceeding the bolus rate while operating in the “PCA” mode. While the software has been revised, the...
Read More
Allergy Laboratories 8/14/98 4. Failure to routinely calibrate, inspect, or check for accuracy and to exercise appropriate controls for automatic, mechanical, or electronic equipment or other types of equipment, including computers, used in the manufacture, processing, packaging, and holding of a drug product according to a written program designed so as to assure performance [21...
Read More
8/12/98 Class II PRODUCT Model No. 250 NeuroCybernetic Prosthesis Programming (NCP) Software, Version 5.0 for use only with the Model No. 100 NCP Pulse Generator and Model No. 200 NCP Programming Wand, an implanted autonomic nerve stimulator implanted for epilepsy. Recall #Z-492-0. REASON The software contains a flaw that reports an inaccurate amount of remaining...
Read More
8/10/98 ACS Centaur Automated Chemiluminescence System 29 units with Software Version 1.0-1.2 The test results will be innacurate if the user changes the reporting units and/or slope and intercept for a test. Z-490-9.
Read More
7/20/98 Hewlett Packard Image Point Ultrasound System Model 2410A used in ultrasound imaging to determine stenosis of the sarotid artery. Software Revisions A.0.0 A.0.1, and A.0.1.2. Class II 441 units. Software overestimates peak blood velocity by 15-25% in some Doppler examinations with the Model 2135A transducer. Recall #=Z-160-8.
Read More
Imaging Sciences International Inc. 7/21/98 CommCat IS-2000 Imaging System dental tomographic and panoramic X-ray machine c) Repair requests beyond the one-year warranty period are not reviewed to determine whether they are complaints, nor is information obtained by the software phone service documented and reviewed for potential complaints.
Read More
Multidata Systems Intl. Corp. 7/16/98 Decision Support Systems (DSS) for Radiation Support device 1. Failure of management with executive responsibility to appoint a member of management to establish authority over and responsibility for (a) ensuring that quality system requirements are effectively established and effectively maintained, and (b) reporting on the performance of the quality system...
Read More
Failure to investigate adverse device effects;failure to install password clearance on computer You failed to investigate the failure of the _____ when operating in MS Access. The system locks up at random and it is unknown whether the software which controls the _____ during _____ which operates off of MS Excel, could be similarly affected....
Read More
Hil-rom 6/25/98 Isolette C2000 Infant Incubators Failure to assure that all inspection and testing equipment is suitable for its intended purposes and is capable of producing valid results. The inspection revealed that the _____ automated simulator test has not been updated to keep current with hardware and software changes made to the C2000 device. (FDA...
Read More
6/25/98 Medtronic 9790C Programmers for Kappa 400 Pacemakekers Model 9952A software: Class II 133 units. Bench testing identified a software anomaly that could result in the delivery of an unitended pacing mode or pacing interval. Recall #=Z-715-8
Read More
6/24/98 Beckman/Coulter Access Immunochemistry Set 81500 Software versions 3.22 and 3.23: Class II 1296 units. Sofware anomaly in these versions may allow test results to be posted to a different patient sample number. Recall #=Z-786/787-8.
Read More
m. concerning the _____ software, used for the handling, storage, and distribution of product…Your response states that the _____ software will be validated. Please note that the establishment of an adequate validation protocol, including system specifications and criteria for demonstrating that the software will meet the requirements, is necessary in order to conduct adequate validation.
Read More
6/17/98 Lifescan Surestep Blood Glucose Monitoring System (Meter of home use) Changed to Class I 727,004 units The meters may give an Er 1 (Error 1)message if a patient’s blood glucose is 500 mg/dL or greater.
Read More
Resuscitaire Radiant Warmer, Infant Radiant Warmer: a) Resuscitaire Radiant Warmer, Model RW81; b) Resuscitaire Radiant Warmer, Model RW82; c) Resuscitaire Radiant Warmer, Model RW82VHA; d) Resuscitaire Radiant Warmer, Model WMRW82; e) Resuscitaire Radiant Warmer, Model WBR 81; f) Resuscitaire Radiant Warmer, Model WBR82. Recall #Z-292/297-9. REASON The device fails to control heater power due to...
Read More
.Select Brand Antacid Peppermint Tablets, 75’s, Lot #7026, manufactured on March 1997, was placed “On Hold”, but was not physically labeled as such, nor did the computerized inventory system indicate this status. Accucal Calcium Antacid Peppermint Tablets, 150’s, Lot #7075 was marked “Rejected” as stored in the warehouse, but this status was not indicated in...
Read More
CIBA Corning ACS 180 Plus Analyzers, used for numerous assays. Recall #Z-359-9. REASON A software error for Troponin I analysis caused erroneous results. CODE All ACS 180 Plus Analyzers with test definition AL software version. MANUFACTURER Chiron Diagnostics Corporation, Oberlin, Ohio. RECALLED BY Manufacturer, by letter/technical bulletin on June 9, 1998. Firm-initiated field correction complete....
Read More
1 54 55 56 57 58

SoftwareCPR Training Courses:

IEC 62304 and other emerging standards for Medical Device and HealthIT Software

Our flagship course for preparing regulatory, quality, engineering, operations, and others for the activities and documentation expected for IEC 62304 conformance and for FDA expectations. The goal is to educate on the intent and purpose so that the participants are able to make informed decisions in the future.  Focus is not simply what the standard says, but what is meant and discuss examples and approaches one might implement to comply.  Special deep discount pricing available to FDA attendees and other regulators.

3-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Brian Pate

Next public offering:  TBD

Email training@softwarecpr.com to request a special pre-registration discount.  Limited number of pre-registration coupons.

Registration Link:

TBD

 


 

Being Agile & Yet Compliant (Public or Private)

Our SoftwareCPR unique approach to incorporating agile and lean engineering to your medical device software process training course is now open for scheduling!

  • Agile principles that align well with medical
  • Backlog management
  • Agile risk management
  • Incremental and iterative software development lifecycle management
  •  Frequent release management
  • And more!

2-days onsite (4 days virtual) with group exercises, quizzes, examples, Q&A.

Instructors: Mike Russell, Ron Baerg

Next public offering: March 7 & 28, 2024

Virtual via Zoom

Registration Link:

Register Now

 


 

Medical Device Cybersecurity (Public or Private)

This course takes a deep dive into the US FDA expectations for cybersecurity activities in the product development process with central focus on the cybersecurity risk analysis process. Overall approach will be tied to relevant standards and FDA guidance documentation. The course will follow the ISO 14971:2019 framework for overall structure but utilize IEC 62304, IEC 81001-5-1, and AAMI TIR57 for specific details regarding cybersecurity planning, risk characterization, threat modeling, and control strategies.

2-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Dr Peter Rech, 2nd instructor (optional)

Next public offering:  TBD

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.