Recipient: Ronald G. Crystal M.D.Product: Investigational biologic products Date: 9/23/02 Subject _____ was enrolled in the study without Computerized Tomography (CT) documentation of the number of _____ present at the time of enrollment. Section 3.2 of the protocol requires that the subject have two or more _____ detected by CT scan. Issuer: Center for Biologics...
Read More
Company: Misys Healthcare Systems IncDate or Enforcement Report: 9/18/02 Class: II PRODUCT Misys Laboratory. Recall # Z-1281-2. REASON Software anomaly regarding cell counts. CODE Versions 5.2, 5.23 and 5.3. RECALLING FIRM/MANUFACTURER Misys Healthcare Systems Inc., Tuscon, AZ, by fax on April 23, 2002. Firm initiated recall is ongoing. VOLUME OF PRODUCT IN COMMERCE 469. DISTRIBUTION...
Read More
Company: Mysis Healthcare Systems, IncDate of Enforcement Report: 9/18/02 Class: II PRODUCT Misys Commercial Laboratory. Recall # Z-1318-2. REASON Software anomaly. If the deformatter begins to process data prior to all patient result records or orders being extracted from the flat files, the remaining results or orders will not be extracted. This can result in...
Read More
Company: Mysis Healthcare Systems, Inc.Date of Enforcement Report: 9/18/02 Class: II PRODUCT Mysis Commercial Laboratory. Recall # Z-1285-2. REASON Software anomaly: a result for an accesion number came across the interface as negative but should have been Do Not Result (DNR). CODE Versions 3.2.0; 3.2.2; 3.2.3, 3.3.0, 3.3.1a, 3.3.1, 3.4.0, 3.4.1b, and 3.4.1a. RECALLING FIRM/MANUFACTURER...
Read More
Company: University of Massachusetts Memorial Medical Center.Product: Blood products Date: 9/12/02 In no more than fifteen (15) working days of receiving this letter, please notify this office in writing of the specific steps you have taken to correct these violations and to prevent them from recurring. Your response should also indicate the status of the...
Read More
/docs/FDAPart11DraftArchivingGuidance00d-1539-nad0001.pdf
Read More
Company: Unit Dose Packaging Inc.Product: Drug products Date: 8/29/02 Failure to establish a written procedure to calibrate, inspect, and check automated, mechanical, or electronic equipment used to manufacture drug products to assure proper performance. [21 CFR 211.68 (a)]. Specifically, there is no data to demonstrate that production equipment (two mixers, one filler, a water system)...
Read More
Company: Diasorin IncDate of Enforcement Report: 8/28/02 Class: II PRODUCT Medusa software, DiaSorin Part 15794, for an OMNI (automated microtiter plate) Instrument used for the ETI-HA-IGMK Plus Enzyme Immunoasay for the Detection of IgM Antibody to Hepatitis A Virus in Human Serum or Plasma. Recall # Z-1265-2. REASON As a result of a software error,...
Read More
Company: Roch DiagnosticsProduct: Accu Check Inform Blood Glucose System Class II Date of Enforcement Report: 8/28/2002 All units that have ever displayed a temperature icon in conjunction with patient use. Reason: May give erroneous results if the temperature icon has ever been displayed with use. Manufacturer: Roch Diagnostics, Mannheim, Germany. Recalled by: Roche Diagnostics Corporation,...
Read More
Company: Dade Behring, IncDate of Enforcement Report: 8/21/02 Class: II PRODUCT Dimension(R) Xpand software revision 6.0.1 and 6.1 (Clinical Chemistry Analyzer). Recall # Z-1226-2. REASON The sample probe does not accurately track the fluid level. CODE Xpand software revision 6.0.1 and 6.1. RECALLING FIRM/MANUFACTURER Recalling Firm: Dade Behring, Inc., Newark, DE, by telephone on May...
Read More
Company: Bayer CorporationDate of Enforcement Report: 8/21/02 Class: III PRODUCT Brand Name: ADVIA Centaur(R) Analyzer. Recall # Z-1233-2. REASON Software programming error. CODE All serial numbers distributed with Model/Part Number 572561, along with all existing software versions from 2.1 to 2.5. RECALLING FIRM/MANUFACTURER Recalling Firm: Bayer Corporation, Business Group Diagnostics, Tarrytown, NY, by letter on...
Read More
Company: University Hospitals of ClevelandProduct: Blood products Date: 8/19/02 With respect to your response to FDA-483 item #2, you stated that you will institute a computer change. We expect that changes to the computer system should be validated. Please provide documentation of the validation for this new change. Additionally, please explain what actions and/or retraining...
Read More
On August 14, 2002, the HIPAA final privacy rule 45 CFR Parts 160 and 164 were modified to respond to comments and to reduce the administrative burden of the rule.  A copy of the new rule can be view here:  HIPAA Modified Final Privacy Rule 2002-08. Medical Device manufacturers that produce devices that will maintain patient...
Read More
Company: Siemens Medical Solutions USADate of Enforcemet Report: 8/14/02 Class: I PRODUCT e.soft software version 2.0 used on all E.CAM Emission Imaging Computed Tomography Systems with e.soft A or e.soft A/P nuclear medicine workstations. Recall # Z-1189-2. REASON Incorrect image time stored for static images resulting in incorrect results. CODE e.soft software version 2.0 installed...
Read More
Company: Baxter Healthcare CorpDate of Enforcement Report: 8/14/02 Class: I PRODUCT Logix-CM Compounder Software. Recall # Z-1203-2. REASON Software allows ingredient concentration errors during compounding. CODE Catalog #2M8400, Logix-CM software version 1.0.2. RECALLING FIRM/MANUFACTURER Baxter Healthcare Corp., Round Lake, IL, by telephone on June 18, 2002. Firm initiated recall is ongoing. VOLUME OF PRODUCT IN...
Read More
http://www.fda.gov/ohrms/dockets/dockets/00d1538/00d1538.htm
Read More
Company: Philips Medical SystemsDate of Enforcement Report: 8/7/02 Class: II PRODUCT 1.5T MR Systems with software version R7.1.2 and higher: Magnetic Resonance Systems are indicated for use as diagnostic devices that produce transverse, sagital, coronal and oblique cross-sectional images, spectroscopic images and/or spectra based upon 1H and 31P metabolites, and that display the internal structure...
Read More
Company: Semens Medical SystemsDate of Enforcement Report: 7/31/02 Class: II PRODUCT SONOLINE Antares, Diagnostic Ultrasound System. The SONOLINE Antares is a high-resolution, digital, broadband diagnostic ultrasound imaging system. It supports the following diagnostic applications: abdominal, OB/GYN, small parts, musculoskeletal, vascular, and urology. Recall # Z-1162-2. REASON Software anomalies that affect measurement and calculation data. CODE...
Read More
Company: Philips Medical Systems Date of Enforcement Report: 7/31/02 Class: II PRODUCT Intera 1.0T and 1.5T MR Systems Magnetic Resonance Systems are indicated for use as diagnostic devices that display the internal structure and/or function of the head, body or extremeties. The coils are used to obtain optimal imaging. Recall # Z-1147-2/Z-1148-2. REASON Higher than...
Read More
Company: Baxter Healthcare CorpDate of Enforcement Report: 7/31/02 Class: II PRODUCT Baxter Ipump Pain Management System infusion pump, product codes 2L3107 and 2L3107R, Baxter Healthcare Corporation, Deerfield, IL. Recall # Z-1166-2. REASON Software anomaly allows access to change prescription or program bolus without security. CODE Product codes 2L3107 and 2L3107R, all serial numbers. RECALLING FIRM/MANUFACTURER...
Read More
Company: Earlham CollegeDate: 07/29/2002 Product: Prenatal vitamins In addition to the above listed violations, our Investigator noted that the laboratory is using an electronic record system for processing and storage of data from the atomic absorption and HPLC instruments that is not set up to control the security and data integrity in that the system...
Read More
Company: Fresenius Medical CareDate: 7/25/02 Product: Naturalyte Acid Concentrate Your firm failed to validate the new salt delivery system for process deviation 2002-005 dated February 1, 2002 for Naturalyte Acid Concentrate lot 2B0017 as required by 21 CFR 820.75. Furthermore, the new computer software included with the new salt delivery system is not validated for...
Read More
Company: Abbott LaboratoriesDate: 7/19/02 Product: Drug products including Biaxin and Meridia FDA compared the MedWatch forms in your computer to your paper files, and this comparison revealed inaccurate data in serious and unexpected adverse drug experience reports submitted to FDA. These inaccurate reports are a violation of 21 CFR 314.80(c)(1)(i). FDA District: Chicago District Office
Read More
Company: Philips Medical SystemsDate of Enforcement Report: 7/3/2002 Class: II PRODUCT Information Center System and Upgrades as follows: a) M3150A Agilent or Virida Information center. Recall # Z-1119-2; b) M3151A Agilent or Virida Information Center Client. Recall # Z-1120-2; c) M3153A Agilent or Virida Surveillance Center. Recall # Z-1121-2; d) M3150AU#A01 Agilent or Virida Information...
Read More
Company: Varian Medical Systems, Inc Date of Enforcement Report: 7/3/2002 Class: II PRODUCT Varian Medical System’s Eclipse VERSION 6.5 software, a proprietary application for Varian Vision ™ imaging systems. Recall # Z-1109-2. REASON Radiation treatment planning software malfunction could result in incorrect treatment regimen. CODE Eclipse software, version 6.5, Serial Numbers: 0016 0018 0022 0034...
Read More
Company: Toshiba America Medical Systems, IncDate of Enforcement Report: 6/26/02 Class: II PRODUCT NEMIO, SSA550A, Diagnostic Ultrasound System. Recall # Z-0923-2. REASON Software anomally caused error in calculating VM_P. CODE Serial Numbers: B1552349 C1572413 C1572414 C1572415 C1572417 C1572418 C1572420 C1572423 C1572428 C1572429 C1572430 C1572431 C1572432 C1572433 C1572421. RECALLING FIRM/MANUFACTURER Toshiba America Medical Systems, Inc., Tustin,...
Read More
483Company: American Red Cross City and Sate: Baltimore, MD FDA District: Baltimore, MD Dates of Inspection: 4/22-6/21/02 4. Failure of the quality control unit to determine root cause and prevent recurrence of significant deviations related to BSD 73.200M, “Shipping” version 1.7 January 2001. Between October 22, 2001 and April 23, 2002, the Chesapeake Region recorded...
Read More
Company: Advanced Sterilization ProductsDate: 6/19/2002 Class: II PRODUCT STERRAD 100S Sterilization System. Recall # Z-1082-2. REASON Software anomaly allows half cycle run when full cycle displayed. CODE Version 3 or lower is subject to upgrade. Newer versions contain software “fix”. Current version is Version 6. RECALLING FIRM/MANUFACTURER Advanced Sterilization Products, Irvine, CA, by letters sent...
Read More
/docs/NEMAPart11DraftValidationGuidanceComments.pdf
Read More
Company: Liebel-Flarsheim CoDate of Enforcement Report: 5/29/2002 Class: II PRODUCT Liebel-Flarsheim Angiomat Illumena Media Power Injectors, P/N 900001, using Version 8.0 software with new console. Recall # Z-1046-2. REASON Pushing the scoll button repeatedly / rapidly results in erroneous protocol ID. CODE Serial Numbers: 0202-5000 to 0402-5126. RECALLING FIRM/MANUFACTURER Recalling Firm: Mallinckrodt, Inc. / Liebel-Flarsheim...
Read More
Company: Fujifilm Medical Systems U.S.ADate of Enforcement Report: 5/29/02 Class: II PRODUCT Synapse Image and Information Management System. Recall # Z-1045-2. REASON Rotated/flipped Images not viewed on the screen may not rotate but left / right markers will be flipped. CODE Synapse Image and Information Management System with software versions 2.0.2 and 2.1.1. RECALLING FIRM/MANUFACTURER...
Read More
http://hissa.nist.gov/effProject/handbook/c++/
Read More
Company: Sigma International MedinaDate of Enforcement Report: 5/8/02 Class: II PRODUCT Sigma 8000 Volumetric Infusion Pumps with software version 3.50.01. Recall # Z-0991-2. REASON Software error: Distal occlusions may not be detected (no alarm) at rates above 550 ml/hr. CODE The following serial numbers. 49105, 49111R, 49188R, 49288R, 49289R, 49402R, 49436R, 49448R, 49454R, 49466R, 49507R,...
Read More
Company: Roche Diagnostics CorpDate of Enforcement Report: 5/1/02 Class: II PRODUCT All DataCare Systems Software, as follows: Recall # Z-0960-2/Z-0974-2. a) DataCare ABG Data Management System for use in a blood gas point of care testing environment; Catalog GD9200; b) DataCare ABG Data Management System for use in a blood gas point of care testing...
Read More
Company: Philips Medical SystemsDate of Enforcement Report: 5/1/2002 Class: III PRODUCT Integris H5000, Angiographic X-ray system. Recall # Z-0924-2. REASON System may occasionally re-boot during a patient examination. CODE Units with SYSCO software version 11.1.2. RECALLING FIRM/MANUFACTURER Recalling Firm: Philips Medical Systems North America, Inc., Bothell, WA, by letter dated November 14, 2001., Manufacturer: Philips...
Read More
Company: Hitachi LtdDate of Enforcement Report: 5/1/02 Class: II PRODUCT Roche Modular Analytics immunoassay based clinical chemistry analyzer, model E170, catalog number 7656961. Recall # Z-0922-2. REASON Possibility of false negative and false positive results being reported. CODE All units with software version 01-03. RECALLING FIRM/MANUFACTURER Recalling Firm: Roche Diagnostics Corp., Indianapolis, IN, by letter...
Read More
Company: Sunquest Information systems, aka Misys HealthcareDate of Enforcement Report: 5/1/02 Class: II PRODUCT Misys Laboratory Microbiology Module, Recall # Z-0916-2. REASON Software does not allow comments to be shown in some reports. CODE Versions 5.2, 5.23 and 5.3. RECALLING FIRM/MANUFACTURER Sunquest Information systems, aka Misys Healthcare, Tucson, AZ. Firm initiated recall is ongoing. VOLUME...
Read More
Company: St. Gobain Ceramiques Avancees DesmarquestProduct: Medical Devices Date: 4/26/02 You need to property validate the … furnace process. Please supply a complete proposed validation protocol for the … furnace. This protocol should specifically address if any aspect of this furnace is computer controlled and how the software will be validated. Enclosed are copies of...
Read More
Company: Zoll Medical CorpDate of Enforcement Report: 4/24/02 Class: I PRODUCT a) Zoll M Series Advisory Defibrillator. Recall # Z-0893-2; b) Zoll M Series AED Defibrillator (semi automatic defibrillator). Recall # Z-0894-2. REASON Defibrillator may fail to detect ventricular fibrillation and fail to deliver shock. CODE a) Serial Numbers: T98F00046-T01K27762 with System Software Version below...
Read More
Company: Sunquest Information Systems, IncDate of Enforcement Report: 4/24/02 Class: II PRODUCT Misys Laboratory for Windows version 5.3. Recall # Z-0888-2. CODE REASON Software QA code of BOTH for patient sex can result in QA flags not being generated for abnormal results. Versions 5.3. RECALLING FIRM/MANUFACTURER Sunquest Information Systems, Inc., aka Misys Healthcare Systems Tucson,...
Read More
A NEMA presentation on HIPAA medical device issues is available here:  NEMA HIPAA Med Dev Issues Presentation. SoftwareCPR® provides on-site and web based training in HIPAA privacy and security regulations, in addition to other regulatory consulting services. SoftwareCPR® also provides a HIPAA Roadmap with links to relevant educational documents to paid subscribers (See Post HIPAA Privacy and...
Read More
A NEMA paper on HIPAA medical device remote service issues is available here: NEMA HIPAA Med Dev Remote Services Paper. SoftwareCPR® provides on-site and web based training in HIPAA privacy and security regulations, in addition to other regulatory consulting services. SoftwareCPR® also provides a HIPAA Roadmap with links to relevant educational documents to paid subscribers (See...
Read More
A NEMA paper on HIPAA is available here:  NEMA HIPAA Security Intro Overview. The Health Insurance Portability and Accountability Act of 1996 (HIPAA) was signed to law on July 21, 1996, and has the general objectives to: Guarantee health insurance coverage of employees Reduce health care fraud and abuse Introduce/implement administrative simplifications in order to augment...
Read More
On December 28, 2000, a final privacy rule 45 CFR Part 160 and 164 was issued.  HHS provides the rule and related guidance here: https://www.hhs.gov/hipaa/for-professionals/privacy/laws-regulations/index.html. Medical device manufacturers that produce devices that will maintain patient data should be aware of HIPAA privacy and security requirements to assure appropriate features are incorporated in their devices to allow...
Read More
Company: American Dental Technologies Inc.Date: 4/18/02 Product: Dental device systems, including PulseMaster 600-IQ and DioLase ST Dental Laser System, PowerPAC High Speed Curing Light System, KCP 5 and 1000 Air Abrasive Kinetic Cavity Preparation System and Ultracam Intraoral Camera System Failure to maintain adequate device master records [21 CFR 820.181] [FDA- 483 Item 5]. For...
Read More
1 80 81 82 83 84 94

SoftwareCPR Training Courses:

IEC 62304 and other emerging standards for Medical Device and HealthIT Software

Our flagship course for preparing regulatory, quality, engineering, operations, and others for the activities and documentation expected for IEC 62304 conformance and for FDA expectations. The goal is to educate on the intent and purpose so that the participants are able to make informed decisions in the future.  Focus is not simply what the standard says, but what is meant and discuss examples and approaches one might implement to comply.  Special deep discount pricing available to FDA attendees and other regulators.

3-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Brian Pate

Next public offering:  June 24-26, 2025 (Boston, MA)

Multiple participants from the same company: If you register 5 or more from the same company before March 15, 2025, receive a special discounted registration of $1999 per person.  These registrations may be transferred to another person at any time. Email training@softwarecpr.com to register and secure the TEAM discount.

 

For private, in-house courses, please contact us.

Email training@softwarecpr.com for more info.

 


 

Being Agile & Yet Compliant (Public)

Our SoftwareCPR unique approach to incorporating agile and lean engineering to your medical device software process training course is now open for registration!

  • Agile principles that align well with medical
  • Backlog management
  • Agile risk management
  • Incremental and iterative software development lifecycle management
  •  Frequent release management
  • And more!

3 days virtual (Zoom) with group exercises, quizzes, examples, Q&A.

Lead Instructor: Mike Russell

Next public offerings:

  • Americas: 11-13 February 2025
  • EU/Eastern Europe/Middle East/Africa/Atlantic/eastern South America: 18-20 February 2025
  • Southern Central Northeastern Pacific: 24-26 February 2025
See our post titled: 1st Quarter 2025 Agile Compliant Courses Scheduled

 

Medical Device Cybersecurity (Public or Private)

This course takes a deep dive into the US FDA expectations for cybersecurity activities in the product development process with central focus on the cybersecurity risk analysis process. Overall approach will be tied to relevant standards and FDA guidance documentation. The course will follow the ISO 14971:2019 framework for overall structure but utilize IEC 62304, IEC 81001-5-1, and AAMI TIR57 for specific details regarding cybersecurity planning, risk characterization, threat modeling, and control strategies.

2-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Dr Peter Rech, 2nd instructor (optional)

Next public offering:  TBD

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.