GE Medical 6/16/96 X-Ray France Failure to validate software programs by adequate and documented testing, when computers are used as part of an automated production or quality assurance system, as required by 21 CFR 820.61. For example, the software control processes have not been validated to correct the most serious faults. Failure to subject any...
Read More
CIBA Corning Acs:180 and ACS:180 Plus Analyzers, used to conduct various laboratory diagnostictests. Recall #Z-368-9. REASON Due to a software error, the ACS Folate and ACS T-Uptake assays may run without the required conditioning solution. As a result, the test results may be outside of the system requirements. CODE All serial numbers for the ACS...
Read More
Somerset Medical Center Blood Bank 5/29/96 1. The validation of the Meditech Computer System is incomplete and/or inadequate for the following reasons: a. Computer documentation did not include testing procedures, system requirements, and system specifications which were specified to Somerset Medical Center. b. There is no assurance that all program changes resulting from the software...
Read More
Planmed OY 5/9/96 Finland 3. Failure to routinely calibrate measurement equipment, to establish adequate calibration procedures, and to maintain records of calibration, including the next calibration date; and failure to validate software programs by adequate and documented testing, when computers are used as part of an automated production or quality assurance system, as required by...
Read More
Cobe 4/16/96 Hemodialysis 10. Failure to have adequate checks designed and implemented to prevent inaccurate data output, input, and programming errors when automated data processing is used for manufacturing or quality assurance purposes, as required by 21 CFR 820.195. For example, the computer search process for complaints by serial number is not designed to detect...
Read More
Abaxis 4/5/96 blood analyzer 1. Failure to validate computer software programs by adequately documented testing, when computers are used as part of an automated production or quality assurance system (21 CFR 820.61). The computer programs which control the automated assembly of rotor parts, loading of reagent beads, dispensing of reagent beads into proper cuvettes, and...
Read More
Barton Carey Medical Products 4/3/96 our Investigator determined that your firm manufactures vascular and burn compression garments. These garments are devices as defined by Section 201(h) of the Federal Food, Drug and Cosmetic Act (the Act). The Inspection revealed that your devices are adulterated in that, the methods used in, or the facilities or controls...
Read More
PIE Medical Equipment 4/1/96 Netherlands ultrasound 1. Failure to maintain records documenting that production and quality assurance measurement equipment is routinely calibrated, inspected, and checked according to written procedures and to validate software programs by adequate and documented testing, when computers are used as part of an automated production or quality assurance system, as required...
Read More
Cabrini Medical Center Blood Bank 2/8/96 14. Failure to maintain written standard operating procedures for the blood bank’s computer system that include: integration of the computer into daily operations; reverting to manual operations when the system is not functioning; control of changes in hardware and software; security procedures to prevent unauthorized access; and maintenance procedures...
Read More
McGHAN Medical Corporation 2/16/96 mammary prostheses The validation of the computerized dip timer system for software version [purged word] and [purged word] fails to assure that the previous files are overwritten when the maximum number of files [purged word] is reached. This is demonstrated by the fact that there was no verification of the file...
Read More
2/14/96 ADAC Laboratories Pinnacle Radiation Therapy Planning Software 83 tapes recalled Due to a software problem, the wedge scatter fields will be calculated incorrectly if the wedge filter is not square. Software version 2.1F
Read More
Ionetics 1/29/96 electrolyte analyzers 1. Failure to ensure specification changes are subject to controls as stringent as those applied to the original specifications of the device [21 CFR 820.100(a)(2)]. For example, your firm failed to document and approve changes made to components and software for your electrolyte analyzers.
Read More
“Information Technology – Software Lifecycle Processes” and issued August 1995. This document was the basis for development of the AAMI Medical Device Software Development Standard and was one of the first software standards recognized by FDA for submission purposes. It can be ordered directly from the IEC at www.iec.ch. Just go to their web store...
Read More
http://www.fda.gov/cdrh/comp/guidance/fod456.pdf
Read More
The paper at the link provided was written by Mark Paulk of the SEI. It compares ISO 9001 and the Capability Maturity Model (CMM). Although there are some differences between ISO 9001 and the FDA Quality System Regulation, both in content and enforcement, this paper may be helpful for medical device software developers involved in...
Read More
Company: Siemens Medical Solutions USA, Inc. Date of Enforcement Report:3/26/2014 Class lI: PRODUCT syngo Dynamics Picture Archiving and Communication System (PACS). Model numbers 10091804, 10091805, 10091806, 10091807, 10091673 Recall Number Z-1216-2014 REASON Potential patient data mixup. Improper error handling could allow for two different patients to be displayed in syngo Dynamics. Images shown for one...
Read More
1 91 92 93

SoftwareCPR Training Courses:

Being Agile & Yet Compliant (Public)

Our SoftwareCPR unique approach to incorporating agile and lean engineering to your medical device software process training course is now open for registration!

  • Agile principles that align well with medical
  • Backlog management
  • Agile risk management
  • Incremental and iterative software development lifecycle management
  •  Frequent release management
  • And more!

3 days virtual (Zoom) with group exercises, quizzes, examples, Q&A.

Lead Instructor: Mike Russell

Next public offerings:

  • Americas: 11-13 February 2025
  • EU/Eastern Europe/Middle East/Africa/Atlantic/eastern South America: 18-20 February 2025
  • Southern Central Northeastern Pacific: 24-26 February 2025
Register using form at this link:     Agile Course Post Promo

 

IEC 62304 and other emerging standards for Medical Device and HealthIT Software

Our flagship course for preparing regulatory, quality, engineering, operations, and others for the activities and documentation expected for IEC 62304 conformance and for FDA expectations. The goal is to educate on the intent and purpose so that the participants are able to make informed decisions in the future.  Focus is not simply what the standard says, but what is meant and discuss examples and approaches one might implement to comply.  Special deep discount pricing available to FDA attendees and other regulators.

3-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Brian Pate

Next public offering:  TBD

Call or email now to schedule a private, in-house class. The fall schedule is filling up!

Email training@softwarecpr.com to request a special pre-registration discount.  Limited number of pre-registration coupons.

Registration Link:

TBD

 


 

Medical Device Cybersecurity (Public or Private)

This course takes a deep dive into the US FDA expectations for cybersecurity activities in the product development process with central focus on the cybersecurity risk analysis process. Overall approach will be tied to relevant standards and FDA guidance documentation. The course will follow the ISO 14971:2019 framework for overall structure but utilize IEC 62304, IEC 81001-5-1, and AAMI TIR57 for specific details regarding cybersecurity planning, risk characterization, threat modeling, and control strategies.

2-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Dr Peter Rech, 2nd instructor (optional)

Next public offering:  TBD

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.