Company:Medtronic Navigation, Inc Date of Enforcement Report 2/11/2009 Class:ll PRODUCT Medtronic O-Arm Intraoperative Imaging System, Catalog number: Bl-700-0027, Recall # Z-0624-2009 REASON Navigation accuracy problem: Accuracy may fall outside of specification when used in combination with the Medtronic Synergy Experience StealthStation System Spine Software with the O-Arm Intraoperative Imaging System and tracker. RECALLING FIRM/MANUFACTURER Recalling...
Read More
Company: Konica Minolta Medical Imaging USA, Inc., Date of Enforcement Report:5/23/2006 Class:lI PRODUCT REGIUS Image Manager (AKA REGIUS IM, RIM). Catalogue Number 5900554. The Konica REGIUS Image Manager Software (RIM) is used to receive and manage computerized X-Ray image file workflow, Recall # Z-0862-06 REASON If the patient demographic data is entered into the CS-1...
Read More
Company:Fischer Imaging Corporation, Date of Enforcement Report:4/19/2006 Class:lI PRODUCT SenoScan Full Field Digital Mammography System, Product number 94001G-3, 105200G, Recall # Z-0757-06 REASON Software upgrade to preclude lost images during mammography procedure. CODE Software versions earlier than 5.2.2 RECALLING FIRM/MANUFACTURER Fischer Imaging Corporation, Denver, CO, by letter on March 27, 2006, firm initiated recall is...
Read More
Company: Konica Medical Imaging, Inc,Date of Enforcement Report: 8/24/05 Class: III PRODUCT Regius-IM, The Regius Image Manager for Centralized workflow and improved image quality software, Recall # Z-1391-05 REASON One feature of the software on some of the devices which permits assembling three digital images into one image has occasionally had problems aligning the three...
Read More
Company: Fischer Imaging CorpDate of Enforcement Report: 7/21/04 Class: II PRODUCT Fischer SenoScan True View Digital Mammography System. Recall # Z-1094-04. REASON Existing software may allow Image data to be truncated in the margin of the breast on processed patient images. CODE Product number 94829G-1, 94830G-1, 94830G-2. RECALLING FIRM/MANUFACTURER Fischer Imaging Corp, Denver, CO, by...
Read More
Company: Imaging Diagnostic Systems Inc.Product: Investigational device Date: 1/28/04 Monitors also failed to observe that the first enrolled patient’s…was assessed with an earlier software version. Because earlier software versions were not equivalent to later versions used in the study, the data should not have been included in the data analysis. Issuer: Center for Devices and...
Read More
Company: Applicare Medical ImagingDate of Enforcement Report: 9/12/01 Class: II RECALL NUMBER, PRODUCT AND CODE: Z-0857-1/Z-0858-1 RadWorks Software Version 5.1 with Software Patch 5 or 7, and RadWorks WebViewer Software Verion 5.1 running on RadWorks 5.1 Software Patch 5 or 7; picture archival and communications system; Applicare Medical Imaging B.V., Zeist, The Netherlands Codes: RadWorks...
Read More
Date of Enforcement Reporl: May 9,2001Class: II PRODUCT: ISS 2001 X-Ray Imaging System. Recall # Z-246-1. REASON: The systems fail to conform to design specifications relating to the emission of electronic product radiation. MANUFACTURER: Fischer Imaging Corporation,Denver, Colorado. RECALLED BY: Manufacturer, to supply new version (1.4) of software by May 1, 2002. Firm- initiated recall...
Read More
Company: Philips Medical Systems Nederland B.V.Date of Enforcement Report: 4/25/01 Date of Recall: February, 2001 Class: II PRODUCT: EasyVision CT/MR Workstation with software release 1.x AND 2.x. Recall #Z-0495-1. REASON: If Ct scans made on different dates, the Image sets are merged into one exam. CODES: EasyVision CT/MR Workstation with release RI.X and release. MANUFACTURER:...
Read More
GE Advantx Vascular Imaging Systems, angiographic X-ray systems: a) GE Advantx Vascular Imaging Systems (DX and DLX) with Stenosis Analysis Software and; b) Cardiac Review Stations (CRS) with Quantitative Analysis Software. Safety Alert #N-002/003-9. REASON When procedures in the Operator Manual for stenosis sizing are not followed each time a time measurement is taken, serious...
Read More
Imaging Sciences International Inc. 7/21/98 CommCat IS-2000 Imaging System dental tomographic and panoramic X-ray machine c) Repair requests beyond the one-year warranty period are not reviewed to determine whether they are complaints, nor is information obtained by the software phone service documented and reviewed for potential complaints.
Read More
Fischer Imaging Corporation 3/19/97 X-ray and mammography systems (a) At least [purged word] systems observed with in-process component and system mis-assemblies such as miswired parts; incorrect or defective printed circuit boards (PCB); missing hardware; defective computer or monitor; incorrect labeling; and defective components such as hand switches, foot switches, collimators, radiation detectors, x-ray tube, cables,...
Read More
Bennett subsidiary of Trex medical Corp. 8/15/96 (component supplier of high frequency programmable generators for use in imaging systems) 2. Validation protocols for software revisions fail to always specify test parameters such as normal input values, boundary values, out-of-range values and the expected outputs values
Read More
The FDA completed its review of an Automated Ejection Fraction SaMD De Novo leading to a new Class II code, ‘QVD.’  The FDA notice follows including the special controls. FDA Notice Re: DEN220063 Trade/Device Name: Caption Interpretation Automated Ejection Fraction Software Regulation Number: 21 CFR 892.2055 Regulation Name: Radiological machine learning-based quantitative imaging software with...
Read More
Company: Shanghai United Imaging Healthcare Co., Ltd Date of Enforcement Report: 4/7/2023 Class II PRODUCT Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography (PET) and Computed Tomography (CT), Model uMI 550 Part Number: 88000657 88000057 Recall Number: Z-1353-2023 REASON: Due to a software issue where the process of patient scanning, the scatter correction may...
Read More
Company: Agfa Healthcare NV Date of Enforcement Report: 1/20/2023 Class II PRODUCT Agfa HealthCare Enterprise Imaging XERO Viewer Recall Number: Z-0775-2023 REASON There is a software defect that can cause issues with images. RECALLING FIRM/MANUFACTURER Agfa Healthcare NV VOLUME OF PRODUCT IN COMMERCE 110 units DISTRIBUTION Distribution throughout US OUS distribution to Australia, Belgium, Brazil,...
Read More
Company: Siemens Medical Solutions USA, Inc Date of Enforcement Report: 4/27/2022 Class II PRODUCT Artis icono biplane- in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use single and biplane diagnostic imaging and interventional procedures. Model: 11327600 Recall Number: Z-0985-2022 REASON Five potential software issues affecting Artis pheno and Artis icono systems...
Read More
Company: Varian Medical Systems Imaging Laboratory GmbH Date of Enforcement Report: 4/20/2022 Class II PRODUCT ARIA Radiation Therapy Management (RTM) VERSIONS 13.6, 15.1, 15.5, 15.6, 16.0, 16.1 Recall Number: Z-0942-2022 REASON Software issue for treatment plan and image management application may result in mismatch values which could result in treatment in the wrong location. RECALLING...
Read More
Company: Siemens Medical Solutions USA, Inc. Date of Enforcement Report: 2/9/2022 Class II PRODUCT ACUSON Sequoia Ultrasound Imaging System with Software Version VA25A/B Recall Number: Z-0547-2022 REASON Due to a calculation error in the measurement when using 2D trace (manual trace) tool. The trace circumference value is overestimated and may potentially result in misdiagnosis of...
Read More
Company: Siemens Medical Solutions USA, Inc. Date of Enforcement Report: 2/2/2022 Class II PRODUCT Artis pheno with software VE20C- diagnostic imaging angiography system Model: 10849000 Recall Number: Z-0515-2022 REASON Four potential software issues with Artis pheno and Artis icono systems with software VE20C: 1. Updated calibration data not saved with measurement after scene+/-; 2. No...
Read More
Company: Philips Healthcare Date of Enforcement Report: 12/1/2021 Class II PRODUCT Philips Azurion systems with software releases 2.1(L1) and 2.1(L2) Recall Number: Z-0238-2022 REASON 3D-RA is a reconstruction software product that can be used with the Philips Azurin system. This product extends the functionality of compatible X-ray equipment with the capability of 3D X-ray imaging....
Read More
Company: Siemens Medical Solutions USA, Inc Date of Enforcement Report: 12/1/2021 Class II PRODUCT Artis zee floor with software VD12-angiography systems developed for single and biplane diagnostic imaging and interventional procedure Model Number: 10094135 Recall Number: Z-0296-2022 Artis zee ceiling with software VD12-angiography systems developed for single and biplane diagnostic imaging and interventional procedure Model...
Read More
Company: Abbott Laboratories Date of Enforcement Report: 12/1/2021 Class II PRODUCT ARCHITECT i1000SR REF 01L86-01/1L86 and 1L87; ARCHITECT i2000SR REF 03M74-02/3M74; ARCHITECT i2000 REF 08C89-01/1G17 and 8C89 Recall Number: Z-0271-2022 ARCHITECT c4000 REF 02P24-01/2P24, 1P86, 1R24, and 1R25; ARCHITECT c8000 REF 01G06-11/1G06; ARCHITECT c16000 REF 03L77-01/3L77; Recall Number: Z-0272-2022 REASON Twelve software-related issues affecting software...
Read More
Company: Philips North America Llc Date of Enforcement Report: 11/17/2021 Class II PRODUCT Philips Azurion Interventional Fluoroscopic X-Ray System Software version: 2.1.x Model numbers: 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228. (Azurion 3 M12, Azurion 3 M15, Azurion 5 M12, Azurion 5 M20, Azurion 7 B12, Azurion 7...
Read More
Company: Baxter Healthcare Corporation Date of Enforcement Report: 11/17/2021 Class II PRODUCT Hemodialysis Delivery System, Software Version 2.x. Recall Number: Z-0215-2022 REASON If the operator initiates therapy with a saved prescription profile and makes a change to the prescription after a disposable filter change using the Same Patient button, the system may suggest values from...
Read More
Company: FujiFilm Healthcare Americas Corporation Date of Enforcement Report: 11/10/2021 Class II PRODUCT Arietta 750 Ultrasound Software Version: V1.0.0 through V2.1.3 Recall Number: Z-0189-2022 Arietta 850 Ultrasound- Software Version: V1.0.0 through V4.1.3 Recall Number: Z-0190-2022 Lisendo 880 Ultrasound Software Version: V1.0 through V4.1.3 Recall Number: Z-0191-2022 Arietta 65 Ultrasound Software Versions: V1.0 through V4.0.0 Recall...
Read More
Company: Siemens Medical Solutions USA, Inc Date of Enforcement Report: 11/10/2021 Class II PRODUCT Artis zeego, Fluoroscopic X-Ray System: Models: 10280959,10502505, 10848283 in combination with the following workplace/system application software models: Product/Model #: syngo X Workplace, 10281061 syngo X Workplace, 10281299 syngo X Workplace, 10502522 Syngo Application Software, 10848815 Recall Number: Z-0200-2022 REASON Software error,...
Read More
Company: Medtronic Neuromodulation Date of Enforcement Report: 11/10/2021 Class II PRODUCT Clinician Programmer Application (CPA) model A610 Clinician Software Application Recall Number: Z-0201-2022 REASON A software anomaly may occur with the clinician programmer application. RECALLING FIRM/MANUFACTURER Medtronic Neuromodulation on 9/23/2021. Voluntary:  Firm Initiated recall is ongoing. VOLUME OF PRODUCT IN COMMERCE 3302 DISTRIBUTION U.S. Nationwide...
Read More
Company: MEDTECH SAS Date of Enforcement Report: 11/3/2021 Class I PRODUCT ROSA One 3.1 Brain application: The device is intended for the spatial positioning and orientation of instruments holders or tool guides to be used by trained surgeons to guide standard surgical instruments during brain and spine surgeries. Recall Number: Z-0118-2022 REASON The firm has...
Read More
Company: AB SCIEX Date of Enforcement Report: 11/3/2021 Class II PRODUCT Cliquid MD (Version 3.4) software used in conjunction with Analyst MD software on SCIEX Citrine & 4500MD series liquid chromatography-tandem mass spectrometry (LC-MS/MS) systems. Model No. 5088288 Recall Number: Z-0175-2022 REASON The values of the Internal Standard (IS) concentrations are incorrectly derived when the...
Read More
Company: Xstrahl Limited Date of Enforcement Report: 11/3/2021 Class II PRODUCT Concerto User Interface Software provided with the following systems: (1) Xstrahl 100 Electronic Brachytherapy – 100kV Superficial X-Ray Therapy System; (2) Xstrahl 150 Electronic Brachytherapy – 150kV Superficial X-Ray Therapy System; (3) Xstrahl 200 – 225kV Superficial / Orthovoltage X-Ray Therapy System; (4) Xstrahl...
Read More
Company: Zimmer Biomet Date of Enforcement Report: 10/29/2021 Class I PRODUCT ROSA One 3.1 Brain Application: The ROSA One 3.1 Brain Application is a robotic platform that helps neurosurgeons with positioning medical instruments or implants during surgery. The device is composed of a compact robotic arm and a touch screen mounted on a stand. Different...
Read More
Company: Medtronic Neuromodulation Date of Enforcement Report: 10/29/2021 Class I PRODUCT ROSA One 3.1 Brain Application: The ROSA One 3.1 Brain Application is a robotic platform that helps neurosurgeons with positioning medical instruments or implants during surgery. The device is composed of a compact robotic arm and a touch screen mounted on a stand. Different...
Read More
Company: Philips Ultrasound Inc Date of Enforcement Report: 10/27/2021 Class II PRODUCT EPIQ CVxi Diagnostic Ultrasound Systems, Product Code: 795232 Recall Number: Z-0123-2022 REASON Ultrasound system software issue can cause an EchoNavigator error notification, leading to an unresponsive system, necessitating system restart, which could lead to therapy/treatment delay and/or unnecessary therapy/treatment. Can occur during ultrasound-guided...
Read More
Company: Canon Medical System, USA, INC. Date of Enforcement Report: 10/27/2021 Class II PRODUCT Canon PET-CT SCANNER, Cartesion Prime, MODEL PCD-1000A Recall Number: Z-0142-2022 REASON A software problem has been identified which could result in the diagnostic imaging system not proceeding to the next actual scan even though automatic start of the next scan is...
Read More
Company: GE Healthcare, LLC Date of Enforcement Report: 10/27/2021 Class II PRODUCT SIGNA Premier Recall Number: Z-0132-2022 SIGNA Architect Recall Number: Z-0133-2022 SIGNA Pioneer Recall Number: Z-0134-2022 Discovery MR750w 3.0T Recall Number: Z-0135-2022 Discovery MR750 3.0T Recall Number: Z-0136-2022 SIGNA Creator, SIGNA Explorer Recall Number: Z-0137-2022 SIGNA Voyager Recall Number: Z-0138-2022 SIGNA Artist Recall Number:...
Read More
Company: Draeger Medical, Inc. Date of Enforcement Report: 10/20/2021 Class II PRODUCT Evita V500 Ventilator with Installed CO2 Measurement Option-Intended for the ventilation of adult, pediatric and neonatal patients. Catalog Number: 8416400 Recall Number: Z-0108-2022 REASON Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of...
Read More
Company: BioMerieux SA Date of Enforcement Report: 10/20/2021 Class II PRODUCT MYLA Versions: 4.7, 4.7.1, 4.8, 4.8.1 and 4.8.2, part of the VIRTUO System Recall Number: Z-0084-2022 REASON Under certain conditions, there is a risk for a false negative result. RECALLING FIRM/MANUFACTURER BioMerieux SA on 9/2/2021. Voluntary:  Firm Initiated recall is ongoing. VOLUME OF PRODUCT...
Read More
Company: Elekta Inc. Date of Enforcement Report: 10/6/2021 Class II PRODUCT Elekta Monaco – Product Usage: used to make treatment plans for patients with prescriptions for external beam radiation therapy. Recall Number: Z-2563-2021 REASON Contour changes can be saved on an unintended image set. In addition, these contour edits do not cause the dose to...
Read More
Company: Biodex Medical Systems, Inc. Date of Enforcement Report: 10/6/2021 Class II PRODUCT Atomlab 500 Dose Calibrator Software 2.0.00 through Revision 2.0.08, inclusive Model: 086-330. Recall Number: Z-2580-2021 REASON When deleting a previously entered custom isotope, the software deletes the isotope, but not the associated dial setting. Possible other stored custom isotopes may then have...
Read More
Company: Philips Ultrasound Inc. Date of Enforcement Report: 9/29/2021 Class II PRODUCT EPIQ Diagnostic Ultrasound Systems: Software Versions: 5.0, 5.0.1 and 5.0.2 Models: EPIQ 5 – 795120, EPIQ 5C – 795205, EPIQ 5G – 795204, EPIQ 7 – 795117, EPIQ 7C – 795201, EPIQ 7G – 795200, EPIQ CVxi -795232, EPIQ CVx – 795231 Recall...
Read More
Company: Philips Ultrasound Inc. Date of Enforcement Report: 9/29/2021 Class II PRODUCT Affiniti Diagnostic Ultrasound Systems: Software Versions: 5.0, 5.0.1 and 5.0.2 Models: Affiniti 30 795121, Affiniti 30 795218, Affiniti 50 795208, Affiniti 50 795118, Affiniti 70 795210, Affiniti 70 795119 Recall Number: Z-2484-2021 REASON Due to a software defect that can intermittently cause the...
Read More
Company: GE Healthcare, LLC Date of Enforcement Report: 9/29/2021 Class II PRODUCT Centricity PACS-IW with Universal Viewer – Product Usage: Is a device that displays medical images (including mammograms) and data from various imaging sources. Recall Number: Z-2502-2021 REASON Image acquisition failures and synchronization failure with the Centricity Enterprise Archive. RECALLING FIRM/MANUFACTURER GE Healthcare, LLC...
Read More
Company: DePuy Orthopaedics, Inc. Date of Enforcement Report: 9/8/2021 Class II PRODUCT TRUMATCH CT PIN GUIDE KIT L- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting Part Number: 420578 Recall...
Read More
Company: Elekta Inc Date of Enforcement Report: 9/8/2021 Class II PRODUCT Monaco RTP System, Software Builds 5.40.00, 5.40.01, 5.40.02, and 5.51.10 Recall Number: Z-2406-2021 REASON If a couch is removed while a plan is loaded, and then added back to the Studyset when the plan is unloaded, it is possible to calculate dose with the...
Read More
Company: Siemens Medical Solutions USA, Inc Date of Enforcement Report: 9/8/2021 Class II PRODUCT Artis Models with SW version VD12 listed below: Artis zee floor 10094135 Artis zee ceiling 10094137 Artis zee MP 10094139 Artis Q floor 10848280 Artis Q ceiling 10848281 Artis Q biplane 10848282 Artis Q.zen floor 10848353 Artis Q.zen ceiling 10848354 Artis...
Read More
Company: Philips Medical Systems Gmbh, DMC Date of Enforcement Report: 9/8/2021 Class II PRODUCT CombiDiagnost R90 Software Version R1.0 and R1.1 Recall Number: Z-2357-2021 REASON Philips has become aware of a software issue, where the kV and mA values are not locked immediately but change when the collimator shutter position is changed directly after the...
Read More
Company: Merge Healthcare, Inc. Date of Enforcement Report: 9/8/2021 Class II PRODUCT Merge Hemo, Software packages 10.2, 10.3, and 10.4 Recall Number: Z-2387-2021 REASON The firm identified a safety issue affecting the software associated with Merge Hemo versions 9.x and 10.x, used in combination with an HL7 lab interface. RECALLING FIRM/MANUFACTURER Merge Healthcare, Inc. on...
Read More
Company: Beckman Coulter Inc. Date of Enforcement Report: 9/1/2021 Class II PRODUCT Normand Remisol Advance Data Manager, REF: C69412, C69413, C44703, and C57017 is a software device that interfaces between laboratory information systems (LIS) and laboratory instruments. Recall Number: Z-2332-2021 REASON There is a potential that the data management system may add additional cells to...
Read More
Company: Philips North America LLC Date of Enforcement Report: 8/25/2021 Class II PRODUCT 728306 Brilliance iCT-Computed Tomography X-ray systems 728305 Brilliance iCT Upgrades Recall Number: Z-2314-2021 REASON When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the...
Read More
1 2 3 4 9

SoftwareCPR Training Courses:

IEC 62304 and other emerging standards for Medical Device and HealthIT Software

Our flagship course for preparing regulatory, quality, engineering, operations, and others for the activities and documentation expected for IEC 62304 conformance and for FDA expectations. The goal is to educate on the intent and purpose so that the participants are able to make informed decisions in the future.  Focus is not simply what the standard says, but what is meant and discuss examples and approaches one might implement to comply.  Special deep discount pricing available to FDA attendees and other regulators.

3-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Brian Pate

Next public offering:  TBD

Email training@softwarecpr.com to request a special pre-registration discount.  Limited number of pre-registration coupons.

Registration Link:

TBD

 


 

Being Agile & Yet Compliant (Public or Private)

Our SoftwareCPR unique approach to incorporating agile and lean engineering to your medical device software process training course is now open for scheduling!

  • Agile principles that align well with medical
  • Backlog management
  • Agile risk management
  • Incremental and iterative software development lifecycle management
  •  Frequent release management
  • And more!

2-days onsite (4 days virtual) with group exercises, quizzes, examples, Q&A.

Instructors: Mike Russell, Ron Baerg

Next public offering: March 7 & 28, 2024

Virtual via Zoom

Registration Link:

Register Now

 


 

Medical Device Cybersecurity (Public or Private)

This course takes a deep dive into the US FDA expectations for cybersecurity activities in the product development process with central focus on the cybersecurity risk analysis process. Overall approach will be tied to relevant standards and FDA guidance documentation. The course will follow the ISO 14971:2019 framework for overall structure but utilize IEC 62304, IEC 81001-5-1, and AAMI TIR57 for specific details regarding cybersecurity planning, risk characterization, threat modeling, and control strategies.

2-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Dr Peter Rech, 2nd instructor (optional)

Next public offering:  TBD