Software Recall

Company: Toshiba America Medical Systems, IncDate of Enforcement Report: 6/26/02 Class: II PRODUCT NEMIO, SSA550A, Diagnostic Ultrasound System. Recall # Z-0923-2. REASON Software anomally caused error in calculating VM_P. CODE Serial Numbers: B1552349 C1572413 C1572414 C1572415 C1572417 C1572418 C1572420 C1572423 C1572428 C1572429 C1572430 C1572431 C1572432 C1572433 C1572421. RECALLING FIRM/MANUFACTURER Toshiba America Medical Systems, Inc., Tustin,...
Read More
Company: Advanced Sterilization ProductsDate: 6/19/2002 Class: II PRODUCT STERRAD 100S Sterilization System. Recall # Z-1082-2. REASON Software anomaly allows half cycle run when full cycle displayed. CODE Version 3 or lower is subject to upgrade. Newer versions contain software “fix”. Current version is Version 6. RECALLING FIRM/MANUFACTURER Advanced Sterilization Products, Irvine, CA, by letters sent...
Read More
Company: Liebel-Flarsheim CoDate of Enforcement Report: 5/29/2002 Class: II PRODUCT Liebel-Flarsheim Angiomat Illumena Media Power Injectors, P/N 900001, using Version 8.0 software with new console. Recall # Z-1046-2. REASON Pushing the scoll button repeatedly / rapidly results in erroneous protocol ID. CODE Serial Numbers: 0202-5000 to 0402-5126. RECALLING FIRM/MANUFACTURER Recalling Firm: Mallinckrodt, Inc. / Liebel-Flarsheim...
Read More
Company: Fujifilm Medical Systems U.S.ADate of Enforcement Report: 5/29/02 Class: II PRODUCT Synapse Image and Information Management System. Recall # Z-1045-2. REASON Rotated/flipped Images not viewed on the screen may not rotate but left / right markers will be flipped. CODE Synapse Image and Information Management System with software versions 2.0.2 and 2.1.1. RECALLING FIRM/MANUFACTURER...
Read More
Company: Sigma International MedinaDate of Enforcement Report: 5/8/02 Class: II PRODUCT Sigma 8000 Volumetric Infusion Pumps with software version 3.50.01. Recall # Z-0991-2. REASON Software error: Distal occlusions may not be detected (no alarm) at rates above 550 ml/hr. CODE The following serial numbers. 49105, 49111R, 49188R, 49288R, 49289R, 49402R, 49436R, 49448R, 49454R, 49466R, 49507R,...
Read More
Company: Hitachi LtdDate of Enforcement Report: 5/1/02 Class: II PRODUCT Roche Modular Analytics immunoassay based clinical chemistry analyzer, model E170, catalog number 7656961. Recall # Z-0922-2. REASON Possibility of false negative and false positive results being reported. CODE All units with software version 01-03. RECALLING FIRM/MANUFACTURER Recalling Firm: Roche Diagnostics Corp., Indianapolis, IN, by letter...
Read More
Company: Sunquest Information systems, aka Misys HealthcareDate of Enforcement Report: 5/1/02 Class: II PRODUCT Misys Laboratory Microbiology Module, Recall # Z-0916-2. REASON Software does not allow comments to be shown in some reports. CODE Versions 5.2, 5.23 and 5.3. RECALLING FIRM/MANUFACTURER Sunquest Information systems, aka Misys Healthcare, Tucson, AZ. Firm initiated recall is ongoing. VOLUME...
Read More
Company: Philips Medical SystemsDate of Enforcement Report: 5/1/2002 Class: III PRODUCT Integris H5000, Angiographic X-ray system. Recall # Z-0924-2. REASON System may occasionally re-boot during a patient examination. CODE Units with SYSCO software version 11.1.2. RECALLING FIRM/MANUFACTURER Recalling Firm: Philips Medical Systems North America, Inc., Bothell, WA, by letter dated November 14, 2001., Manufacturer: Philips...
Read More
Company: Roche Diagnostics CorpDate of Enforcement Report: 5/1/02 Class: II PRODUCT All DataCare Systems Software, as follows: Recall # Z-0960-2/Z-0974-2. a) DataCare ABG Data Management System for use in a blood gas point of care testing environment; Catalog GD9200; b) DataCare ABG Data Management System for use in a blood gas point of care testing...
Read More
Company: Sunquest Information Systems, IncDate of Enforcement Report: 4/24/02 Class: II PRODUCT Misys Laboratory for Windows version 5.3. Recall # Z-0888-2. CODE REASON Software QA code of BOTH for patient sex can result in QA flags not being generated for abnormal results. Versions 5.3. RECALLING FIRM/MANUFACTURER Sunquest Information Systems, Inc., aka Misys Healthcare Systems Tucson,...
Read More
Company: Zoll Medical CorpDate of Enforcement Report: 4/24/02 Class: I PRODUCT a) Zoll M Series Advisory Defibrillator. Recall # Z-0893-2; b) Zoll M Series AED Defibrillator (semi automatic defibrillator). Recall # Z-0894-2. REASON Defibrillator may fail to detect ventricular fibrillation and fail to deliver shock. CODE a) Serial Numbers: T98F00046-T01K27762 with System Software Version below...
Read More
Company: Misys Healthcare SystemsDate of Enforcement Report: 4/17/02 Class: II PRODUCT Misys Laboratory (formerly FlexiLab) information system. Recall # Z-0803-2. REASON Software anomaly may allow patient results to be transposed against another patient header or identification. CODE Versions 5.2, 5.23, and 5.3. RECALLING FIRM/MANUFACTURER Misys Healthcare Systems, Tuscon, AZ, by notice on March 8, 2002....
Read More
Class: III Aeroset Chloestorol Reagent Lot 82099HW00. Firm’s list no. 7D62-01. Manufacturer: Abbott Laboratories, Inc., Sout Pasadena, CA Recalled by: Manufacturer, by letters on April 17, 2002. Firm initiated recall is ongoing Distribution: Nationwide and internatinally; 715 units were distributed Reason: Stability may be compromised over time, product currently OK Recall number: Z-1031-2
Read More
Company: Sunquest Information Systems, IncDate of Enforcement Report: 4/10/02 Class: II PRODUCT FlexiLab Laboratory Information System. Recall # Z-0710-2. REASON Date Format Issue–Flexilab Software converts specimen collect and receipt dates to American standard format ‘mm/dd/yyyy’. For sites using the Euro date format this can alter the generation of QA warnings or flags based on collect...
Read More
Company: Roche Diagnostics CorpDate of Enforcement Report: 4/10/02 Class: II PRODUCT All DataCare Systems Software, as follows: a) DataCare ABG Data Management System for use in a blood gas point of care testing environment; Catalog GD9200 b) DataCare ABG Data Management System for use in a blood gas point of care testing environment; Catalog GD9201...
Read More
Company: Abbott Laboratories, Inc Date of Enforcement Report: 3/24/02 Class: II PRODUCT AEROSET System Software v1.00ER005.2 and v1.0ER005 Catalog Numbers: 2-8996-02 and 2-89996-01. Recall # Z-0594-2/0595-2. REASON System may incorrectly associate sample ID & test requests undercertain conditions of non-bar coded or unreadable bar coded samples. CODE All Codes are affected by this field correction....
Read More
Company: Siemens AG Medical EngineeringDate of Enforcement Report: 3/27/02 Class: III PRODUCT a) ACOM.PC Angiographic X-ray System; b) Digital Image Communications System. Recall # Z-0603-2. REASON Software problem in recording heights and weights. CODE Siemens part number: a) ACOM.PC V3.0-4371105 b) Quantcor V4.0-4786898, 5915553, 5915579. Equipment Concerned: Product: Part No. Serial No. LVA V 4.0...
Read More
Company: Bayer CorporationDate of Enforcement Report: 3/27/02 Class: III PRODUCT Clinitek 100 Software Kit. Recall # Z-0634-2. REASON False negative results when high concentrations of ketone exist. CODE Version 7.00. RECALLING FIRM/MANUFACTURER Recalling Firm: Bayer Corp., Elkhart, IN, by telephone and fax on January 21, 2002. Manufacturer: Bayer Corporation, Mishawaka, IN. Firm initiated recall is...
Read More
Company: Sunquest Information Systems IncDate of Enforcement Report: 3/6/02 Class: II PRODUCT FlexiLab Laboratory Information System, Version 5.2, 5.23, and 5.3, Recall # Z-0531-02. REASON Software glitch can cause results to be filed under an incorrect accession number under certain conditions. CODE Versions 5.2, 5.23, and 5.3. RECALLING FIRM/MANUFACTURER Sunquest Information Systems Inc., Tucson, AZ,...
Read More
Company: ADAC LaboratoriesDate of Enforcement Report: 2/13/02 Class: II PRODUCT ADAC brand Pinnacle Radiation Therapy Planning System Radiation Therapy Treatment Software version 5.2g, Recall # Z-0524-02. REASON A Software problem that could result in incorrect radiation dosage being applied. CODE Model: 9200-0579A. RECALLING FIRM/MANUFACTURER ADAC Laboratories, Milpitas, CA, by letters on Sept. 26, 2000. Firm...
Read More
Company: Sunquest Information Systems, IncDate of Enforcement Report: 2/13/02 Class: II PRODUCT FlexiLab? Laboratory Information System, Recall # B-0710-2. REASON Blood bank computer system, which had a software glitch or defect, was distributed. CODE Versions 5.23 and 5.3 with RBR module. RECALLING FIRM/MANUFACTURER Sunquest Information Systems, Inc., Tucson, AZ, by letter dated March 19, 2001....
Read More
Company: Sunquest Information Systems, Inc Date of Enforcement Report: 2/13/02 Class: II PRODUCT FlexiLab? Blood Bank and Blood Donor Software, Recall # B-0711-2. REASON Blood bank computer system, which had a software glitch or defect, was distributed. CODE Versions 5.2 and 5.23. RECALLING FIRM/MANUFACTURER Sunquest Information Systems, Inc., Tucson, AZ, by letter dated June 22,...
Read More
Company: Sunquest Information Systems, IncDate of Enforcement Report: 2/13/2002 Class: II PRODUCT FlexiLab? Blood Bank and Blood Donor Software, Recall # B-0709-2. REASON Blood bank computer system, which had a software glitch or defect, was distributed. CODE Versions 5.2, 5.23 and 5.3. RECALLING FIRM/MANUFACTURER Sunquest Information Systems, Inc., Tucson, AZ, by letter dated March 13,...
Read More
Company: Roche Diagnostics CorpDate of Enforcement Report: 2/6/02 Class: II PRODUCT All DataCare Systems Software, as follows: A) DataCare Server/Workstation Computer for DataCare ABG Data Management System for use in a blood gas point of care testing environment; Catalog GD9200. Sold in the U.S. B) DataCare Networked Client Computer for DataCare ABG Data Management System...
Read More
Company: Mediware Information Systems, IncDate of Enforcement Report: 2/30/02 Class: III PRODUCT Hemocare Blood Bank Data Management System Software, Recall # B-0626-2. REASON Computer software that could allow the assignment of an extended expiration date to blood products was distributed. CODE Versions 5.1, 5.2, 5.2a, and 5.2b. RECALLING FIRM/MANUFACTURER Mediware Information Systems, Inc., Melville, NY...
Read More
Company: Dade Behring, IncDate of Enforcement Report: 1/30/02 Class: II PRODUCT Dimension Clinical Chemical Analyzer XL/RxL/ARx with software revision 5.2, Recall # Z-0433-02. REASON Instructions to software revision are missing a step. CODE Revision 5.2 The following serial numbers are involved in the recall: 971807, 973002, 970457, 973278, 970598, 970454, 970049, 971979, 973242, 970864, 973218,...
Read More
Company: Sunquest Information Systems, IncDate of Enforcement Report: 1/23/02 Class: II PRODUCT FlexiLab? Blood Bank and Blood Donor Software (3 versions), Recall # B-0627-2. REASON Blood establishment computer software that could allow the release of unsuitable blood products was distributed. CODE Version numbers 5.2, 5.23, and 5.3. RECALLING FIRM/MANUFACTURER Sunquest Information Systems, Inc., Tucson, AZ,...
Read More
Company: Biochem Immunosystems (U.S.), IncDate of Enforcement Report: 1/23/02 Class: II PRODUCT Personal Lab Version 2.2a. The device is an Immunochemistry Analyzer Open System, Recall # Z-0463-02. REASON Samples were not picked up and added to the dilution tube. CODE Version 2.2a Catalog # PL01. RECALLING FIRM/MANUFACTURER Recalling Firm: Biochem Immunosystems (U.S.), Inc. Allentown, PA,...
Read More
Company: Mediware Information Systems, IncDate of Enforcement Report: 1/23/02 Class: II PRODUCT Hemocare Blood Bank Data Management System, Recall # B-0670-2. REASON Defects in the design of blood bank software programs could result in the release of unsuitable blood products. CODE Version 5.1, 5.2 and/or 5.2a. RECALLING FIRM/MANUFACTURER Mediware Information Systems, Inc., Melville, NY, by...
Read More
Company: Mediware Information Systems, IncDate of Enforcement Report: 1/23/02 Class: II PRODUCT Hemocare Blood Bank Data Management System, Recall # B-0668-2. REASON Defects in the design of blood bank software programs could result in the release of unsuitable blood products. CODE Version 5.1, 5.2 and/or 5.2a. RECALLING FIRM/MANUFACTURER Mediware Information Systems, Inc., Melville, NY, by...
Read More
Company: Mediware Information Systems, IncDate of Enforcement Report: 1/23/02 Class: 111 PRODUCT Hemocare Blood Bank Data Management System, Recall # B-0671-2. REASON Defects in the design of blood bank software programs could result in the release of unsuitable blood products. CODE Version 5.1, 5.2 and/or 5.2a. RECALLING FIRM/MANUFACTURER Mediware Information Systems, Inc., Melville, NY, by...
Read More
Company: Mediware Information Systems, Inc.Date of Enforcement Report: 1/23/02 Class: III PRODUCT Hemocare Blood Bank Data Management System, Recall # B-0669-2. REASON Defects in the design of blood bank software programs could result in the release of unsuitable blood products. CODE Version 5.1, 5.2 and/or 5.2a. RECALLING FIRM/MANUFACTURER Mediware Information Systems, Inc., Melville, NY, by...
Read More
Company: Mediware Information Systems, Inc.Date of Enforcement Report: 1/23/02 Class: III PRODUCT Hemocare Blood Bank Data Management System, Recall # B-0667-2. REASON Defects in the design of blood bank software programs could result in the release of unsuitable blood products. CODE Version 5.1, 5.2 and/or 5.2a. RECALLING FIRM/MANUFACTURER Mediware Information Systems, Inc., Melville, NY, by...
Read More
Company: Mediware Information Systems, IncDate of Enforcement Report: 1/23/02 Class: III PRODUCT Hemocare Blood Bank Data Management System, Recall # B-0666-2. REASON Defects in the design of blood bank software programs could result in the release of unsuitable blood products. CODE Version 5.1, 5.2 and/or 5.2a. RECALLING FIRM/MANUFACTURER Mediware Information Systems, Inc., Melville, NY, by...
Read More
Company: Siemens Medical Systems, IncDate of Enforcement Report: 1/16/02 Class: III PRODUCT Autoperfusion Software used on the ICON workstations designed for use with the E.CAM Emission Imaging Computed Tomography Systems to conduct automated myocardial processing in diagnosing myocardial disease in cardiac SPECT studies. Recall # Z-0427-02. REASON The software did not save oblique data correctly...
Read More
Company: Siemens Medical Systems, INCDate of Enforcement Report: 11/16/01 Class: III PRODUCT Autoperfusion Software, which includes The Cardiac Review (Profile Review) application, used on the ICON workstations designed for use with the E.CAM Emission Imaging Computed Tomography Systems to conduct automated myocardial processing in diagnosing myocardial disease in cardiac SPECT studies. Recall # Z-0428-02. REASON...
Read More
Company: Abbott Laboratories, IncDate of Enforcement Report: 11/28/01 Class: II PRODUCT The Abbott AxSYM System is a fully automated Immunoassay analyzer designed to perform Microparticle Enzyme Immunoassay (MEIA), Fluorescence Polarization Immunoassay (FPIA), Radiative Energy Attenuation (REA) Assay Technology, and Ion Capture (IC) Immunoassay Technologies. The AxSYM System performs random access, continuous access, and STAT processing...
Read More
Company: Multidata Systems International Corp.Date of Enforcement Report: 11/28/01 Class: II PRODUCT Radiation Treatment Planning Decision Support System (DSS), Recall # Z-0213-2. REASON The software calculates radiation overdoses that require human intervention to detect and prevent injury to radiation therapy patients. CODE All versions. RECALLING FIRM/MANUFACTURER Multidata Systems International Corp., St. Louis, MO, by letter...
Read More
Company: Radionics, A Division of Tyco HealthcareDate of Enforcement Report: 11/21/01 Class: II PRODUCT Radionics XKnife RT Software, Recall # Z-0275-2. REASON Software error may cause incorrect jaw settings for a record/verify system. CODE Version 1.02 and lower. RECALLING FIRM/ MANFACTURER Radionics, A Division of Tyco Healthcare, Billerica, MA, by telephonebetween 9/17/2001 and 18/01. Firm...
Read More
Company: Novametrix Medical Systems, IncDate of Enforcement Report: 11/7/01 Class: II RECALL NUMBER, PRODUCT AND CODE: Z-0216-2, Novametrix NICO Monitor, Model 7300 with software versions below 4.3 REASON: Monitor may miscalculate Pulse Rate and Oxygen Saturation (SpO2) values. MANUFACTURER/RECALLING FIRM: Novametrix Medical Systems, Inc., Wallingford, CT RECALLED BY: Recalling firm by letter on 7/9/01 FIRM...
Read More
Company: Marconi Medical Systems, IncClass: II Date of Enforcement Report: 10/31/01 RECALL NUMBER, PRODUCT AND CODE: Z-0181-2, Marconi MX8000 Computed Tomography X-ray System, Codes: All Marconi MX8000 CT Scanners, running with Version 2.0 and 2.01 software. REASON: Software in version 2.0 / 2.01 software may lead to improper patient treatment. MANUFACTURER/RECALLING FIRM: Marconi Medical Systems,...
Read More
Company: Nihon Kohden, IncDate of Enforcement Report: 10/24/01 Class: II RECALL NUMBER, PRODUCT AND CODE: Z-0177-2/Z-0178-2, Procyon Bedside Monitors, Models BSM-4104A and BSM- 4114A. Codes: Models BSM-4104A and BSM-4114A Bedside Monitors. REASON: Software causes infusion rate 2.2 times proper rate. MANUFACTURER/RECALLING FIRM: Nihon Kohden, Inc., CA RECALLED BY: Recalling firm by telephone and letter on...
Read More
Company: Siemens Medical Systems, IncDate of Enforcement Report: 10/24/01 Class: III RECALL NUMBER, PRODUCT AND CODE: Z-0031-2, Motion Electronics Computer Chassis (MECC) Backplanes installed in E.CAM Emission Imaging Computed Tomography Systems used for all common nuclear medicine procedures; Siemens Medical Systems, Inc., Nuclear Medicine Group, Hoffman Estates, IL Codes: MECC Backplane “2104938 REV C” —...
Read More
Company: Dade Behring Marburg GMBHDate of Enforcement Report: 10/24/01 Class: II RECALL NUMBER, PRODUCT AND CODE: Z-0013-2, Behring Coagulation System (BCS) Coagulation Analyzer. Codes: Software Version 2.2 Code OVIO03 REASON: Falsely shortened APTT value. MANUFACTURER/RECALLING FIRM: Dade Behring Marburg GMBH, Marburg, Germany RECALLED BY: Dade Behring, Inc., Newark, DE, by letter on 6/25/01 FIRM INITIATED...
Read More
Company: Marconi Medical Systems, IncDate of Enforcement Report: 10/24/01 Class: III RECALL NUMBER, PRODUCT AND CODE: Z-0057-2, Marconi MX8000 Computed Tomography X-ray Codes: All Marconi MX8000 CT Scanners running with Version 2.0 sofware. REASON: A software anomaly in ver 2.0 software may result in improper patient diagnosis. MANUFACTURER/RECALLING FIRM: Marconi Medical Systems, Inc., Highland Heights,...
Read More
Company: TegimentaDate of Enforcement Report: 10/17/01 Class: II RECALL NUMBER, PRODUCT AND CODE: Z-0008-2, COBAS Integra 400 laboratory clinical chemistry analyzer, Model 400, Catalog number 1045199. All serial numbers of analyzers lower than 38-2606. Serial numbers of 38-2606 or higher have already been updated with software revision upgrade 4. REASON: If the operator fails to...
Read More
Company: Bayer CorporationDate of Enforcement Report: 10/10/01 Class: III RECALL NUMBER, PRODUCT AND CODE: Z-1223-1, Software versions AE through AL (except AI) for the Bayer Advia Centaur Automated Chemiluminescence Immunoassay System. The affected software was distributed as a component of the Advia Centaur Automated Chemiluminescence Immunoassay system. Codes: Software versions AE through AL, except version...
Read More
Company: SIEMENS MEDICAL SYSTEMS, INCDate of Enforcement Report: 9/19/01 Class: II RECALL NUMBER, PRODUCT AND CODE: Z-1106-1 – Z-1108-1 Programmable Multi-Axis Controller (PMAC 2 Lite) circuit board installed in E.CAM Emission Imaging Computed Tomography Systems used for all common nuclear medicine procedures; Codes: a) Model 4380213 – E.CAM Gantry Single Head, serial numbers 0001 through...
Read More
Compnay: Siemens Medical Systems, IncDate of Enforcement Report: 9/12/01 Class: II RECALL NUMBER, PRODUCT AND CODE: Z-0851-1 ICON Profile Processing Software used on the ICON P workstations designed for use with E.CAM Gamma Camera Systems to conduct cardiac Tomo and SPECT applications; 60195-5203. Codes: All E.CAM gamma cameras with the Profile option with ICON-P workstations...
Read More
Company: Applicare Medical ImagingDate of Enforcement Report: 9/12/01 Class: II RECALL NUMBER, PRODUCT AND CODE: Z-0857-1/Z-0858-1 RadWorks Software Version 5.1 with Software Patch 5 or 7, and RadWorks WebViewer Software Verion 5.1 running on RadWorks 5.1 Software Patch 5 or 7; picture archival and communications system; Applicare Medical Imaging B.V., Zeist, The Netherlands Codes: RadWorks...
Read More
1 58 59 60 61 62 63

SoftwareCPR Training Courses:

Being Agile & Yet Compliant (Public)

Our SoftwareCPR unique approach to incorporating agile and lean engineering to your medical device software process training course is now open for registration!

  • Agile principles that align well with medical
  • Backlog management
  • Agile risk management
  • Incremental and iterative software development lifecycle management
  •  Frequent release management
  • And more!

3 days virtual (Zoom) with group exercises, quizzes, examples, Q&A.

Lead Instructor: Mike Russell

Next public offerings:

  • Americas: 11-13 February 2025
  • EU/Eastern Europe/Middle East/Africa/Atlantic/eastern South America: 18-20 February 2025
  • Southern Central Northeastern Pacific: 24-26 February 2025
See our post titled: 1st Quarter 2025 Agile Compliant Courses Scheduled

 

IEC 62304 and other emerging standards for Medical Device and HealthIT Software

Our flagship course for preparing regulatory, quality, engineering, operations, and others for the activities and documentation expected for IEC 62304 conformance and for FDA expectations. The goal is to educate on the intent and purpose so that the participants are able to make informed decisions in the future.  Focus is not simply what the standard says, but what is meant and discuss examples and approaches one might implement to comply.  Special deep discount pricing available to FDA attendees and other regulators.

3-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Brian Pate

Next public offering:  June 24-26, 2025 (Boston, MA)

Special pre-registration discount through January 31, 2025. Group discounts available. See Post on this course.

For private, in-house courses, please contact us.

Email training@softwarecpr.com for more info.

 


 

Medical Device Cybersecurity (Public or Private)

This course takes a deep dive into the US FDA expectations for cybersecurity activities in the product development process with central focus on the cybersecurity risk analysis process. Overall approach will be tied to relevant standards and FDA guidance documentation. The course will follow the ISO 14971:2019 framework for overall structure but utilize IEC 62304, IEC 81001-5-1, and AAMI TIR57 for specific details regarding cybersecurity planning, risk characterization, threat modeling, and control strategies.

2-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Dr Peter Rech, 2nd instructor (optional)

Next public offering:  TBD

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.