Tag

recall

Software Recall

4/6/99 Class II Phillips Medical 1086 units. Integris family of x-ray controls and generators (fluoroscopy,, urology, cardiology). The defect occurs when the system is driven to maximum EER and the source to image receptor distance (SID) is moved toa shorter distance while continuing to make exposures. In this manner of operation, the output may exceed...
Read More
3/29/99 Digisonics Class II OBLink software releases prior to 9.0 39 units. Y2K sofvtware error may produce an incorrect estimated delivery date (EDD). Z-1056-9.
Read More
Eagle 3000 Floorloader Steam Sterilizer. Recall #Z-936-9. REASON There is an error in the control system software, which causes the chamber temperatures to exceed the upper limit. CODE Various serial numbers. MANUFACTURER Steris Corporation, Erie Pennsylvania. RECALLED BY Steris Corporation, Mentor, Ohio, by Customer Notifications on March 10, 1999. Firm-initiated recall ongoing. DISTRIBUTION Nationwide, Canada,...
Read More
3/5/99 Class II Hewlett Packard 145 units. Software used with transducer EV7014 incorrectly indicates track of biospy needle. Z-772-9
Read More
Philips Medical Sonodiagnost 800, intended for endovaginal biopsy for various gynecologic and obstetric procedures including follicular retrieval, endometrial biopsy, and fluid aspiration of any type in a potentially pregnant patient. Recall #Z-772-9. REASON Software used with Transducer EV7014 incorrectly indicates track of the biopsy needle. CODE All SD800’s with Software versions 2.1. and 2.1.1 used...
Read More
All IMMULITE 2000 Automated Immunoassay Analyzers with SoftwareVersion 1.1, a clinical automated immunoassay analyzer intended for in-vitro diagnostic use for detection of analytates in human specimens. Recall #Z-909-9. REASON If the initial counts of the reaction tubes are exactly 10,000 cps, this leads to erroneously low calculated results. CODE All serial numbers in commercial distribution...
Read More
Abbott Cell-Dyn 3000 Series Hematology Systems, Cell-Dyn Reticulocyte Reagent, when used in combination with Cell-Dyn 3500 reticulocyte software enables a whole blood specimen to be analyzed for reticulocytes. The Reticulocyte Reagent is intended for in vitro diagnostic use (outside the body). Recall #Z-715-9. REASON These lots may exhibit elevated Reticulocyte results. In some cases, the...
Read More
2/16/99 Class II 128XP Acuson 637 units. Rev 27 software with OB calculations options are affected. Potential error condition related to OB reports printed after diagnosis. Z-787-9
Read More
Solar 7000 and Solar 8000 Patient Monitors with Version 5D Software, used to display physiological data from modules which monitor the patient for ECG, blood pressure, cardiac output, respiration, pulseoximetry, etc. Recall #Z-727/728-9. REASON A software deviation can affect the alarm default settings, which may result in incorrect or missed alarm notification. CODE Solar 7000/8000...
Read More
10-Feb -99 Precision Therapy International. 190 units. Class II. 3-D Imgcomp Utility User Manual Update. Version 2.72b & 3.07d. A design error may leadto a serious miscalculation of radiation dose. Z-1223-9
Read More
2/9/99 Class II Bayer AADVIA120 Hematology System 920 units. The device may give erroneous reslts because of the valve gasket deterioration in the Baso channel check valve, and the software designed will not flag significant discrepancies in the results between the two methodologies at extremely low counts.Z-869/870-9
Read More
2/5/99 Class I recall Marquette Electronics ECG, blood pressure, cariac output, respiraion, pulse oximetry Monitors. 607 new units and 97 update kits. A software deviation can affect alarm default settings, which may result in incorrect or missed alarm notification. Z-727/728-9
Read More
2/1/99 Abbott Alcyon 300 Analyzer (IVD) Possible mixup of patient results. 631 units. Z-617-9
Read More
National Biomedical Computer System, Versions 1.1 to 1.3.2. Recall #B-861-9. REASON Y2K defect in the computer software incorrectly calculates the end date of a temporary deferral to be 1900 instead of 2000, and removes the deferral from the database. CODE Software versions 1.1 to 1.3.2. MANUFACTURER American Red Cross Blood Services, Arlington, Virginia. RECALLED BY...
Read More
Four Recalls of Picker International Nuclear Cameras and imaging systems12/15/98 Prism XPVT 63 units. Unexpected camera movement which requires the operator to power down and reboot with delay in diagnosis. Z-643/650-9 11/18/98 Model 3000 325 units. Unintended motion of gantry during setup. Z-641/642-9 11/17/98 ModelFX AND fx 800 WITH VERSION 8.3.11 software. 70 units. Software...
Read More
12/10/98 Datalink Management System Software Versions 4.1-4.2 Beckman Coulter 248 units. The devices may allow the test results from an associated clinical instrument to be placed under an incorrect sample ID. Z-829-9
Read More
PRODUCTVersion 6A and prior versions of software for Marquette Coherent Digital Telemetry (CDT) LAN Monitoring Systems, a patient monitoring system designed to collect and transmit ECG and other physiological data from ambulatory patients, without the patient being physically connected to a display device. Recall #Z-774-0.
Read More
11/24/98 HMT Medical Technologies Switzerland. Lithotripsy instruments Software versions 3.0 and 1.2. 17 units. Software upgrade yielding error messages and complete lock-up of instruments. Z-680/681
Read More
11/2/98 ACS Centaur Automated Chemiluminescence System 244 units with Software Version 1.2 The device may go into premature ready state with bleach still remaining in the analyzeer lines due to a software error. Z-499-9.
Read More
Software for the Abbott AxSYM System, a fully automated immunoanalyzer. Recall #Z-331-9. REASON Extra assay results from a previous run can appear with new sample results, as well as error messages, due to software defect in the internal result database management. CODE All software versions below 3.03. MANUFACTURER Abbott Laboratories, Irving, Texas. RECALLED BY Abbott...
Read More
9/4/98 Nera Electronics/Baxter Healthcare Flo Gard 8200 Volumetric Infusion Pump: Class II 13 units. Board assemblies had the wrong software version installed on them. The boards have version 2.09 instead of version 2.13. Recall #=-Z-777-8.
Read More
10/1/98 Class II Quinton Q-Cath Hemodynamic Analysis System 81 units. Units with version 4.5 software. Device has potential to display previous patient’s blood pressure. Z-855/859-9
Read More
GE Advantx Vascular Imaging Systems, angiographic X-ray systems: a) GE Advantx Vascular Imaging Systems (DX and DLX) with Stenosis Analysis Software and; b) Cardiac Review Stations (CRS) with Quantitative Analysis Software. Safety Alert #N-002/003-9. REASON When procedures in the Operator Manual for stenosis sizing are not followed each time a time measurement is taken, serious...
Read More
Blood Bank Control Systems. Recall #B-020-0. REASON Computer software contains programming errors that could potentially result in the release of a unit that has not been tested for antibodies. CODE Software versions 4.0 and greater. MANUFACTURER Blood Bank Computer Systems, Kent, Washington. RECALLED BY Manufacturer, by letter dated September 1, 1998. Firm-initiated field correction complete....
Read More
8/12/98 Class II PRODUCT Model No. 250 NeuroCybernetic Prosthesis Programming (NCP) Software, Version 5.0 for use only with the Model No. 100 NCP Pulse Generator and Model No. 200 NCP Programming Wand, an implanted autonomic nerve stimulator implanted for epilepsy. Recall #Z-492-0. REASON The software contains a flaw that reports an inaccurate amount of remaining...
Read More
8/10/98 ACS Centaur Automated Chemiluminescence System 29 units with Software Version 1.0-1.2 The test results will be innacurate if the user changes the reporting units and/or slope and intercept for a test. Z-490-9.
Read More
7/20/98 Hewlett Packard Image Point Ultrasound System Model 2410A used in ultrasound imaging to determine stenosis of the sarotid artery. Software Revisions A.0.0 A.0.1, and A.0.1.2. Class II 441 units. Software overestimates peak blood velocity by 15-25% in some Doppler examinations with the Model 2135A transducer. Recall #=Z-160-8.
Read More
6/25/98 Medtronic 9790C Programmers for Kappa 400 Pacemakekers Model 9952A software: Class II 133 units. Bench testing identified a software anomaly that could result in the delivery of an unitended pacing mode or pacing interval. Recall #=Z-715-8
Read More
6/24/98 Beckman/Coulter Access Immunochemistry Set 81500 Software versions 3.22 and 3.23: Class II 1296 units. Sofware anomaly in these versions may allow test results to be posted to a different patient sample number. Recall #=Z-786/787-8.
Read More
6/17/98 Lifescan Surestep Blood Glucose Monitoring System (Meter of home use) Changed to Class I 727,004 units The meters may give an Er 1 (Error 1)message if a patient’s blood glucose is 500 mg/dL or greater.
Read More
Resuscitaire Radiant Warmer, Infant Radiant Warmer: a) Resuscitaire Radiant Warmer, Model RW81; b) Resuscitaire Radiant Warmer, Model RW82; c) Resuscitaire Radiant Warmer, Model RW82VHA; d) Resuscitaire Radiant Warmer, Model WMRW82; e) Resuscitaire Radiant Warmer, Model WBR 81; f) Resuscitaire Radiant Warmer, Model WBR82. Recall #Z-292/297-9. REASON The device fails to control heater power due to...
Read More
CIBA Corning ACS 180 Plus Analyzers, used for numerous assays. Recall #Z-359-9. REASON A software error for Troponin I analysis caused erroneous results. CODE All ACS 180 Plus Analyzers with test definition AL software version. MANUFACTURER Chiron Diagnostics Corporation, Oberlin, Ohio. RECALLED BY Manufacturer, by letter/technical bulletin on June 9, 1998. Firm-initiated field correction complete....
Read More
Venturi Ventilator, provides a software that continuously adaptsto the patient’s changing breathing demands and assures ventilation at the lowest pressure. Recall #Z-803-9. REASON Software error may cause ventilator to deliver more than set tidal volume. CODE Venturi Ventilator with Software Revision C. Serial #970222. MANUFACTURER Cardiopulmonary Corporation, Milford, Connecticut. RECALLED BY Manufacturer, by letter on...
Read More
Information Management System for Mammography Practices. Recall #Z-187-9. REASON The software was manufactured with methods that resulted in defective information being provided to the physician, and therefore, the patients with positive mammograms may not follow-up with a biopsy exam. CODE All copies of Mammoworks Version 2.6. MANUFACTURER Lumisys, Inc., (formerly Compurad), Tucson, Arizona. RECALLED BY...
Read More
ACS:Centaur Automated Chemiluminescence System, used to conduct various laboratory diagnostictests. Recall #Z-363-9. REASON Software errors for V1.0 which could lead to erroneous test results. CODE All lot numbers for the ACS Centaur. MANUFACTURER Chiron Diagnostics Corporation, Oberlin, Ohio. RECALLED BY Manufacturer, by fax dated March 25, 1998. Firm-initiated field correction complete. DISTRIBUTION Germany and Italy....
Read More
3/17/98 Recon Rx neurological biofeedback monitoring1762 units Rx unit were not functioning correctly. Obsolete version 3.01 software had been installed in the units.
Read More
Venturi Ventilator, used to provide a software that continouslyadapts to the patient’s changing breathing demands and assures ventilation at the lowest pressure. The display monitor provides pertient patient parameters and waveforms. Recall #Z-802-9. REASON Display monitor may go blank due to faulty inverter. CODE Serial numbers: 970192, 970190, 970154, 970164, 970169, 970172, 970138, 970161, 970163,...
Read More
CIBA Corning ACS:180 SE Analyzer, used to conduct various laboratory diagnostic tests. Recall#Z-369-9. REASON Software error causes 16 different assays to not calculate SI units correctly. CODE All lot numbers for ACS 180 SE Analyzers. MANUFACTURER Chiron Diagnostics Corporation, Oberlin, Ohio. RECALLED BY Manufacturer, by fax on November 25, 1997. Firm-initiated field correction complete. DISTRIBUTION...
Read More
9/30/97 EZ-Flow 480 Ambulatory Infusion Pump 1058 devices. Defective software may cause over infusion when using the PCA mode. 1/9/98 confirmed a patient death due to a software logic error related to use in the patient controlled analgesia (PCA) mode caused the lethal dose of morphine sulfate. Two non-fatal incidents in 1996 also occurred.
Read More
7/7/97 Solar 7000/8000 Patient Monitors with Software Revision 4A783 units. A software defect is causing incorrect waveform data.
Read More
McGaw Stratus Ambulatory Infusion System Model 550 and Alphamed EZ Pump, Model 550 withSoftware Version 2.32.00/2.42.01. Recall #Z-188-9. REASON The pump could under infuse or blood could back up into IV lines when the pumps are running at a slow flow rate with a large infusion set. CODE Serial Numbers: P61145, P71108, P71115, P71125, P71130,...
Read More
6/5/97 HP M1026a Anesthetic Gas Analyzer with O2 sensor2,284 units. The units may display innacurrate oxygen reading due to software problem.
Read More
3/13/97 Sulzer Intermedics Marathon SR & DR Pacemakers 3300 units RX 5000 Software Operating System A software condition caused the parameters of the 2 pacemakers to become reversed. That is, the software cannot distinguish Marathon DR (dual chamber) from Marathon SR (single-chamber) USER PRESET values; whichever values were most recently stored will be recalled.
Read More
1/10/97 Quinton Instrument Stress Test Monitors – 3 models ~7000units affected.A software anomaly could cause their treadmills to erroneously increase or decrease in speed and/or grade when the “Stop EXER” key is pressed at the end of a stage.
Read More
9/27/96 Radiation Oncology Computer Systems (ROCS) Treatment Planning System Software Version 5.0.x. An error occurs when using this software version for brachytherapy dose estimations for user specified calculation points for re-oriented linear sources.
Read More
CIBA Corning Acs:180 and ACS:180 Plus Analyzers, used to conduct various laboratory diagnostictests. Recall #Z-368-9. REASON Due to a software error, the ACS Folate and ACS T-Uptake assays may run without the required conditioning solution. As a result, the test results may be outside of the system requirements. CODE All serial numbers for the ACS...
Read More
2/14/96 ADAC Laboratories Pinnacle Radiation Therapy Planning Software 83 tapes recalled Due to a software problem, the wedge scatter fields will be calculated incorrectly if the wedge filter is not square. Software version 2.1F
Read More
Company: Siemens Medical Solutions USA, Inc. Date of Enforcement Report:3/26/2014 Class lI: PRODUCT syngo Dynamics Picture Archiving and Communication System (PACS). Model numbers 10091804, 10091805, 10091806, 10091807, 10091673 Recall Number Z-1216-2014 REASON Potential patient data mixup. Improper error handling could allow for two different patients to be displayed in syngo Dynamics. Images shown for one...
Read More
1 61 62 63

SoftwareCPR Training Courses:

IEC 62304 and other emerging standards for Medical Device and HealthIT Software

Our flagship course for preparing regulatory, quality, engineering, operations, and others for the activities and documentation expected for IEC 62304 conformance and for FDA expectations. The goal is to educate on the intent and purpose so that the participants are able to make informed decisions in the future.  Focus is not simply what the standard says, but what is meant and discuss examples and approaches one might implement to comply.  Special deep discount pricing available to FDA attendees and other regulators.

3-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Brian Pate

Next public offering:  June 24-26, 2025 (Boston, MA)

Multiple participants from the same company: If you register 5 or more from the same company before March 15, 2025, receive a special discounted registration of $1999 per person.  These registrations may be transferred to another person at any time. Email training@softwarecpr.com to register and secure the TEAM discount.

 

For private, in-house courses, please contact us.

Email training@softwarecpr.com for more info.

 


 

Being Agile & Yet Compliant (Public)

Our SoftwareCPR unique approach to incorporating agile and lean engineering to your medical device software process training course is now open for registration!

  • Agile principles that align well with medical
  • Backlog management
  • Agile risk management
  • Incremental and iterative software development lifecycle management
  •  Frequent release management
  • And more!

3 days virtual (Zoom) with group exercises, quizzes, examples, Q&A.

Lead Instructor: Mike Russell

Next public offerings:

  • Americas: 11-13 February 2025
  • EU/Eastern Europe/Middle East/Africa/Atlantic/eastern South America: 18-20 February 2025
  • Southern Central Northeastern Pacific: 24-26 February 2025
See our post titled: 1st Quarter 2025 Agile Compliant Courses Scheduled

 

Medical Device Cybersecurity (Public or Private)

This course takes a deep dive into the US FDA expectations for cybersecurity activities in the product development process with central focus on the cybersecurity risk analysis process. Overall approach will be tied to relevant standards and FDA guidance documentation. The course will follow the ISO 14971:2019 framework for overall structure but utilize IEC 62304, IEC 81001-5-1, and AAMI TIR57 for specific details regarding cybersecurity planning, risk characterization, threat modeling, and control strategies.

2-days onsite with group exercises, quizzes, examples, Q&A.

Instructor: Dr Peter Rech, 2nd instructor (optional)

Next public offering:  TBD

Corporate Office

15148 Springview St.
Tampa, FL 33624
USA
+1-781-721-2921
Partners located in the US (CA, FL, MA, MN, TX) and Canada.